Searched for: +
(1 - 3 of 3)
document
Smit, G. (author), Plettenburg, D.H. (author), Van der Helm, F.C.T. (author)
Rejection rates of upper limb prostheses are high (23-45%). Amputees indicate that the highest design priority should be reduction of the mass of the prosthetic device. Despite all efforts, the mass of the new prosthetic hands is 35-73% higher than that of older hands. Furthermore current hands are thicker than a human hand, they operate slower...
journal article 2014
document
Smit, G. (author), Plettenbrug, D.H. (author), Van der Helm, F.C.T. (author)
Background: Cosmetic gloves that cover a prosthetic hand have a parasitic positive stiffness that counteracts the flexion of a finger joint. Objectives: Reducing the required input torque to move a finger of a prosthetic hand by compensating the parasitic stiffness of the cosmetic glove. Study design: Experimental, test bench. Methods: The...
journal article 2014
document
Smit, G. (author), Plettenburg, D.H. (author), Van der Helm, F.C. (author)
The goal of this study was to find an efficient method of energy transmission for application in an anthropomorphic underactuated body-powered (BP) prosthetic hand. A pulley-cable finger and a hydraulic cylinder finger were designed and tested to compare the pulley-cable transmission principle with the hydraulic cylinder transmission principle....
journal article 2014