Searched for: author%3A%22Tabacchini%2C+V.%22
(1 - 1 of 1)
document
Borghi, G. (author), Tabacchini, V. (author), Bakker, P.J. (author), Schaart, D.R. (author)
Recently, a monolithic scintillator detector for time-of-flight (TOF)/depth-of-interaction (DOI) positron emission tomography (PET) was developed. It has a detector spatial resolution of ∼1.7 mm full-width-at-half-maximum (FWHM), a coincidence resolving time (CRT) of ∼215 ps FWHM, and ∼4.7 mm FWHM DOI resolution. Here, we demonstrate, for the...
journal article 2018