Xiao, Yidi (author), Rivandi, H. (author), Lopes Marta da Costa, T.M. (author)Emerging ultrasound (US) biomedical applications, from battery-powered US imaging to US neuromodulation, demand wearable form factor and power-efficient US transmitters. Fulfilling these specifications demands a high-frequency and power-efficient 2D US phased-array transmitter directly integrated with the ASIC. In such systems, pulsers are the...conference paper 2023
- document
-
Hopf, Y.M. (author), Ossenkoppele, B.W. (author), Soozande, M. (author), Noothout, E.C. (author), Chang, Z.Y. (author), Vos, H.J. (author), Bosch, Johan G. (author), Verweij, M.D. (author), de Jong, N. (author), Pertijs, M.A.P. (author)In this letter, a compact high-voltage (HV) transmit circuit for dense 2-D transducer arrays used in 3-D ultrasonic imaging systems is presented. Stringent area requirements are addressed by a unipolar pulser with embedded transmit/receive switch. Combined with a capacitive HV level shifter, it forms the ultrasonic HV...journal article 2022
- document
-
Chen, C. (author), Pertijs, M.A.P. (author)As medical ultrasound imaging moves from conventional cart-based scanners to new form factors such as imaging catheters, hand-held point-of-care scanners and ultrasound patches, there is an increasing need for integrated transceivers that can be closely integrated with the transducer to provide channel-count reduction, improved signal quality...journal article 2021