Measuring handrim wheelchair propulsion in the lab

A critical analysis of stationary ergometers

Journal Article (2020)
Author(s)

Rick De Klerk (University Medical Center Groningen)

Riemer Jakobs K. Vegter (University Medical Center Groningen)

Vicky L. Goosey-Tolfrey (Loughborough University)

Barry S. Mason (Loughborough University)

John P. Lenton (British Cycling)

Dirkjan H.E.J. Veeger (TU Delft - Biomechanical Engineering, Vrije Universiteit Amsterdam)

L.H.V. van der Woude (University Medical Center Groningen)

Department
Biomechanical Engineering
DOI related publication
https://doi.org/10.1109/RBME.2019.2942763
More Info
expand_more
Publication Year
2020
Language
English
Department
Biomechanical Engineering
Volume number
13
Pages (from-to)
199-211
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

There are many ways to simulate handrim wheelchair propulsion in the laboratory. Ideally, these would be able to, at least mechanically, simulate field conditions. This narrative review provides an overview of the lab-based equipment used in published research and critically assesses their ability to simulate and measure wheelchair propulsion performance. A close connection to the field can only be achieved if the instrument can adequately simulate frictional losses and inertia of real-life handrim wheelchair propulsion, while maintaining the ergonomic properties of the wheelchair-user interface. Lab-based testing is either performed on a treadmill or a wheelchair ergometer (WCE). For this study WCEs were divided into three categories: roller, flywheel, and integrated ergometers. In general, treadmills are mechanically realistic, but cannot simulate air drag and acceleration tasks cannot be performed; roller ergometers allow the use of the personal wheelchair, but calibration can be troublesome; flywheel ergometers can be built with commercially-available parts, but inertia is fixed and the personal wheelchair cannot be used; integrated ergometers do not employ the personal wheelchair, but are suited for the implementation of different simulation models and detailed measurements. Lab-based equipment is heterogeneous and there appears to be little consensus on how to simulate field conditions.