Using Acoustic Vibrations as a Method for Implant Insertion Assessment in Total Hip Arthroplasty

Journal Article (2022)
Author(s)

J.C. Wei (TU Delft - Medical Instruments & Bio-Inspired Technology)

W.H.A. Crezee (Student TU Delft)

H. Jongeneel (Student TU Delft)

T.S.A. de Haas (Student TU Delft)

W.L.A. Kool (Student TU Delft)

Bryan J. Blaauw (Elkerliek Hospital)

Jenny Dankelman (TU Delft - Medical Instruments & Bio-Inspired Technology)

T. Horeman (TU Delft - Medical Instruments & Bio-Inspired Technology)

Research Group
Medical Instruments & Bio-Inspired Technology
Copyright
© 2022 J.C. Wei, W.H.A. Crezee, H. Jongeneel, T.S.A. de Haas, W.L.A. Kool, Bryan J. Blaauw, J. Dankelman, T. Horeman
DOI related publication
https://doi.org/10.3390/s22041609
More Info
expand_more
Publication Year
2022
Language
English
Copyright
© 2022 J.C. Wei, W.H.A. Crezee, H. Jongeneel, T.S.A. de Haas, W.L.A. Kool, Bryan J. Blaauw, J. Dankelman, T. Horeman
Research Group
Medical Instruments & Bio-Inspired Technology
Issue number
4
Volume number
22
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

The success of total hip arthroplasty depends on the experience of the surgeon, and one of the ways the surgeon currently determines the final implant insertion depth is to listen to the change in audible pitch of the hammering sound. We investigated the use of vibration emissions as a novel method for insertion quality assessment. A non-invasive contact microphone-based measurement system for insertion depth estimation, fixation and fracture detection was developed using a simplified in vitro bone/implant (n = 5). A total of 2583 audio recordings were analyzed in vitro to obtain energy spectral density functions. Out of the four main resonant peaks under in vitro conditions, broach insertion depth statistically correlates to increasing 3rd and 4th peak frequencies. Degree of fixation was also observed as higher goodness of fit (0.26–0.78 vs. 0.12–0.51 between two broach sizes, the latter undersized). Finally, however, the moment of fracture could not be predicted. A cadaveric in situ pilot study suggests comparable resonant frequencies in the same order of magnitudes with the bone model. Further understanding of the signal patterns are needed for an early warning system diagnostic system for imminent fractures, bone damage, improving accuracy and quality of future procedures