Hyperpolarised gas filling station for medical imaging using polarised 129Xe and 3He

Journal Article (2021)
Author(s)

Wai Tung Lee (Australia Nuclear Science and Technology Organization, New South Wales 2234)

Gang Zheng (Monash University)

Cavin L. Talbot (University of Queensland)

Xin Tong (Spallation Neutron Source Science Center, Dongguan , Institute of High Energy Physics Chinese Academy of Science)

Tim D'Adam (Australia Nuclear Science and Technology Organization, New South Wales 2234)

Steven R. Parnell (TU Delft - RID/TS/Instrumenten groep)

Michael de Veer (Monash University)

Graham Jenkin (Monash University and Hudson Institute of Medical Research)

Graeme R. Polglase (Monash University, Monash University and Hudson Institute of Medical Research)

undefined More Authors (External organisation)

Research Group
RID/TS/Instrumenten groep
DOI related publication
https://doi.org/10.1016/j.mri.2021.02.010 Final published version
More Info
expand_more
Publication Year
2021
Language
English
Research Group
RID/TS/Instrumenten groep
Volume number
79
Pages (from-to)
112-120
Page Views
322
Collections
Institutional Repository
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

We report the design, construction, and initial tests of a hyperpolariser to produce polarised 129Xe and 3He gas for medical imaging of the lung. The hyperpolariser uses the Spin-Exchange Optical Pumping method to polarise the nuclear spins of the isotopic gas. Batch mode operation was chosen for the design to produce polarised 129Xe and polarised 3He. Two-side pumping, electrical heating and a piston to transfer the polarised gas were some of the implemented techniques that are not commonly used in hyperpolariser designs. We have carried out magnetic resonance imaging experiments demonstrating that the 3He and 129Xe polarisation reached were sufficient for imaging, in particular for in vivo lung imaging using 129Xe. Further improvements to the hyperpolariser have also been discussed.

Files

Lee_2021_Hyperpolarised_Gas_Fi... (pdf)
(pdf | 0.806 Mb)
- Embargo expired in 26-03-2022