Reproducibility of the Pleth Variability Index in premature infants

Journal Article (2017)
Author(s)

Wilhelmina J. Den Boogert (Erasmus MC)

Hugo A. Van Elteren (Erasmus MC)

T. G. Goos (TU Delft - Medical Instruments & Bio-Inspired Technology, Erasmus MC)

Irwin Reiss (Erasmus MC)

Rogier C.J. de Jonge (Erasmus MC)

Victor J. van Den Berg (Erasmus MC)

Research Group
Medical Instruments & Bio-Inspired Technology
Copyright
© 2017 Wilhelmina J. Den Boogert, Hugo A. Van Elteren, T.G. Goos, Irwin K.M. Reiss, Rogier C.J. De Jonge, Victor J. van Den Berg
DOI related publication
https://doi.org/10.1007/s10877-017-0058-3
More Info
expand_more
Publication Year
2017
Language
English
Copyright
© 2017 Wilhelmina J. Den Boogert, Hugo A. Van Elteren, T.G. Goos, Irwin K.M. Reiss, Rogier C.J. De Jonge, Victor J. van Den Berg
Research Group
Medical Instruments & Bio-Inspired Technology
Issue number
3
Volume number
32 (June 2018)
Pages (from-to)
457-464
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

The aim was to assess the reproducibility of the Pleth Variability Index (PVI), developed for non-invasive monitoring of peripheral perfusion, in preterm neonates below 32 weeks of gestational age. Three PVI measurements were consecutively performed in stable, comfortable preterm neonates in the first 48 h of life. On each occasion, pulse oximeter sensors were attached to two different limbs for 5 min. Reproducibility was assessed with the intra-class correlation coefficient (ICC) and Bland–Altman analysis. A total of 25 preterm neonates were included. Inter-limb comparison showed fair to moderate ICC’s with 95%-confidence intervals (95%-CI). Left hand–right hand ICC = 0.498, 95%-CI (0.119–0.753); right foot–right hand ICC = 0.314 (−0.088–0.644); right foot–left foot ICC = 0.315 (−0.089–0.628). Intra-limb comparison showed fair to moderate ICC for right foot–right foot ICC = 0.380 (−0.014–0.677); and good ICC for right hand–right hand ICC = 0.646 (0.194–0.852). Bland–Altman plots showed moderate reproducibility of measurements between different limbs and of the same limb in consecutive time periods, with large biases and wide limits of agreement. The findings from this study indicate that PVI measurement is poorly reproducible when measured on different limbs and on the same limb in stable and comfortable preterm neonates.