Estimated ECG Subtraction method for removing ECG artifacts in esophageal recordings of diaphragm EMG

Journal Article (2021)
Author(s)

Annemijn H. Jonkman (Amsterdam UMC)

Ricardo Juffermans (Amsterdam UMC)

Jonne Doorduin (Radboud University Medical Center)

Leo M.A. Heunks (Amsterdam UMC)

Jaap Harlaar (Erasmus MC, TU Delft - Biomechatronics & Human-Machine Control)

Research Group
Biomechatronics & Human-Machine Control
Copyright
© 2021 Annemijn H. Jonkman, Ricardo Juffermans, Jonne Doorduin, Leo M.A. Heunks, J. Harlaar
DOI related publication
https://doi.org/10.1016/j.bspc.2021.102861
More Info
expand_more
Publication Year
2021
Language
English
Copyright
© 2021 Annemijn H. Jonkman, Ricardo Juffermans, Jonne Doorduin, Leo M.A. Heunks, J. Harlaar
Research Group
Biomechatronics & Human-Machine Control
Bibliographical Note
Green Open Access added to TU Delft Institutional Repository 'You share, we take care!' - Taverne project https://www.openaccess.nl/en/you-share-we-take-care Otherwise as indicated in the copyright section: the publisher is the copyright holder of this work and the author uses the Dutch legislation to make this work public.@en
Volume number
69
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

The accuracy of diaphragm electromyogram (EMGdi) derived parameters, as used in critically ill intensive care unit (ICU) patients, can be compromised due to electrocardiographic (ECG) interference in the EMGdi signal. Removal of ECG contamination from the esophageal recordings of the EMGdi is challenging due to spectral overlapping of EMG and ECG signals and because of variability in ECG shape and amplitude. Therefore, we designed an Estimated ECG Subtraction (EES) method, based on three steps: (1) identification of the timing of the ECG artifact without an ECG reference channel, (2) estimation of the normalized ECG, considering the EMGdi as noise, and (3) subtraction of the denormalized ECG estimate from the EMGdi recordings. We evaluated the EES method against the use of a single wavelet-based adaptive filter. Using EMGdi signals of ten ICU patients and simulated contaminated EMG, we demonstrated that the EES method yields uncontaminated EMGdi, and showed that it is more effective than a wavelet-based adaptive filter only. Implementation of this technique may offer means to improve diaphragm activity monitoring and control in clinical practice.