Blood-flow Volume Estimation by a 2-D Sparse Array

Journal Article (2025)
Author(s)

Claudio Giangrossi (University of Florence)

Alessandro Ramalli (University of Florence)

Francesco Guidi (University of Florence)

E.C. Noothout (TU Delft - ImPhys/Verweij group)

Luxi Wei (Erasmus MC)

H.J. Vos (Erasmus MC, TU Delft - ImPhys/Verweij group)

P. Tortoli (University of Florence)

Research Group
ImPhys/Verweij group
DOI related publication
https://doi.org/10.1016/j.ultrasmedbio.2025.06.005
More Info
expand_more
Publication Year
2025
Language
English
Research Group
ImPhys/Verweij group
Issue number
9
Volume number
51
Pages (from-to)
1580-1588
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

Objective: The assessment of blood-flow volume (BFV) is clinically relevant for the diagnosis and monitoring of cardiovascular dysfunctions and the prevention of subsequent secondary diseases. Non-invasive BFV measurement based on ultrasound methods are appealing for lower cost, real-time operation, and equipment portability. Recently, complex ultrasound research scanners with 1024 channels controlling the elements of a 2-D matrix array probe, have been demonstrated suitable for off-line accurate BFV estimates. In this work, a streamlined approach, using a 256-channel research scanner paired with a 256-element 2-D sparse spiral array, is proposed and validated. Methods: This setup allows for simultaneous scanning of the vessel's longitudinal and transverse sections through an interleaved transmission sequence. In real-time, the longitudinal scan is used to determine the flow direction, while the transverse scan captures both the dynamic cross-sectional area and the local velocities by high frame rate color flow mapping. Results: Flow phantom experiments under steady and pulsatile flow conditions were conducted to assess the performance by comparing the measurements with the outputs of a reference flow sensor. The proposed method provided accurate and precise BFV values for both flow conditions, with mean percentage error and standard deviation always lower than 9.4% and 2.8%, respectively. Furthermore, preliminary in vivo experiments have produced results consistent with those reported in the literature. Conclusion: The proposed method based on the use of a sparse array has permitted accurate and precise phantom BFV measurements and has been shown suitable for real-time arterial BFV measurements.