Cost-efficient anthropomorphic head phantom for quantitative image quality assessment in cone beam CT

Journal Article (2022)
Author(s)

Yichao Wang (Philips Healthcare Nederland)

Jenny Dankelman (TU Delft - Medical Instruments & Bio-Inspired Technology)

Danny Ruijters (Philips Healthcare Nederland)

DOI related publication
https://doi.org/10.1088/2057-1976/aca02d Final published version
More Info
expand_more
Publication Year
2022
Language
English
Bibliographical Note
Green Open Access added to TU Delft Institutional Repository 'You share, we take care!' - Taverne project https://www.openaccess.nl/en/you-share-we-take-care Otherwise as indicated in the copyright section: the publisher is the copyright holder of this work and the author uses the Dutch legislation to make this work public.
Journal title
Biomedical Physics and Engineering Express
Issue number
6
Volume number
8
Article number
065038
Downloads counter
296
Collections
Institutional Repository
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

In this study, a novel anthropomorphic head phantom for quantitative image quality assessment in cone beam computed tomography (CBCT) is proposed. The phantom is composed of tissue equivalent materials (TEMs) which are suitable for cost-efficient fabrication methods such as silicone casting and 3D printing. A monocalcium phosphate/gypsum mixture (MCPHG), nylon and a silyl modified polymer gel (SMP) are proposed as bone, muscle and brain equivalent materials respectively. The TEMs were evaluated for their radiodensity in terms of Hounsfield Units (HU) and their x-ray scatter characteristics. The median radiodensity and inter quartile range (IQR) of the MCPHG and SMP were found to be within the range of the theoretical radiodensity for bone and brain tissue: 922 (IQR = 156) and 47 (IQR = 7) HU respectively. The median radiodensity of nylon was slightly outside of the HU range of muscle tissue, but within the HU range of a combination of muscle and adipose tissue: −18 (IQR = 40) HU. The median ratios between the measured scatter characteristics and simulated tissues were between 0.84 and 1.13 (IQR between 0.05 and 0.14). The preliminary results of this study show that the proposed design and TEMs are potentially suitable for the fabrication of a cost-efficient anthropomorphic head phantom for quantitative image quality assessment in CT or CBCT.