The recently developed versatile emission computed tomography (VECTor) technology enables high-energy SPECT and simultaneous SPECT and PET of small animals at sub-mm resolutions. VECTor uses dedicated clustered pinhole collimators mounted in a scanner with three stationary large-area NaI(Tl) gamma detectors. Here, we develop and validate dedicated image reconstruction methods that compensate for image degradation by incorporating accurate models for the transport of high-energy annihilation gamma photons. Ray tracing software was used to calculate photon transport through the collimator structures and into the gamma detector. Input to this code are several geometric parameters estimated from system calibration with a scanning 99mTc point source. Effects on reconstructed images of (i) modelling variable depth-of-interaction (DOI) in the detector, (ii) incorporating photon paths that go through multiple pinholes ('multiple-pinhole paths' (MPP)), and (iii) including various amounts of point spread function (PSF) tail were evaluated. Imaging 18F in resolution and uniformity phantoms showed that including large parts of PSFs is essential to obtain good contrast-noise characteristics and that DOI modelling is highly effective in removing deformations of small structures, together leading to 0.75 mm resolution PET images of a hot-rod Derenzo phantom. Moreover, MPP modelling reduced the level of background noise. These improvements were also clearly visible in mouse images. Performance of VECTor can thus be significantly improved by accurately modelling annihilation gamma photon transport. ...
Introduction
High-resolution pre-clinical 131I SPECT can facilitate development of new radioiodine therapies for cancer. To this end, it is important to limit resolution-degrading effects of pinhole edge penetration by the high-energy γ-photons of iodine. Here we introduce, optimize and validate 131I SPECT performed with a dedicated high-energy clustered multi-pinhole collimator.

Methods
A SPECT–CT system (VECTor/CT) with stationary gamma-detectors was equipped with a tungsten collimator with clustered pinholes. Images were reconstructed with pixel-based OSEM, using a dedicated 131I system matrix that models the distance- and energy-dependent resolution and sensitivity of each pinhole, as well as the intrinsic detector blurring and variable depth of interaction in the detector. The system performance was characterized with phantoms and in vivo static and dynamic 131I-NaI scans of mice.

Results
Reconstructed image resolution reached 0.6 mm, while quantitative accuracy measured with a 131I filled syringe reaches an accuracy of + 3.6 ± 3.5% of the gold standard value. In vivo mice scans illustrated a clear shape of the thyroid and biodistribution of 131I within the animal. Pharmacokinetics of 131I was assessed with 15-s time frames from the sequence of dynamic images and time–activity curves of 131I-NaI.

Conclusions
High-resolution quantitative and fast dynamic 131I SPECT in mice is possible by means of a high-energy collimator and optimized system modeling. This enables analysis of 131I uptake even within small organs in mice, which can be highly valuable for development and optimization of targeted cancer therapies. ...
SPECT with submegabecquerel amounts of tracer or subsecond time resolution would enable a wide range of new imaging protocols such as screening tracers with initially low yield or labeling efficiency, imaging low receptor densities, or even performing SPECT outside regular radiation laboratories. To this end we developed dedicated ultra-high-sensitivity pinhole SPECT.
METHODS:
A cylindric collimator with 54 focused 2.0-mm-diameter conical pinholes was manufactured and mounted in a stationary small-animal SPECT system. The system matrix for image reconstruction was calculated via a hybrid method based on both (99m)Tc point source measurements and ray-tracing analytic modeling. SPECT images were reconstructed using pixel-based ordered-subsets expectation maximization. Performance was evaluated with phantoms and low-dose bone, dynamic kidney, and cardiac mouse scans.
RESULTS:
The peak sensitivity reached 1.3% (13,080 cps/MBq). The reconstructed spatial resolution (rod visibility in a micro-Jaszczak phantom) was 0.85 mm. Even with only a quarter megabecquerel of activity, 30-min bone SPECT scans provided surprisingly high levels of detail. Dynamic dual-isotope kidney and (99m)Tc-sestamibi cardiac scans were acquired with a time-frame resolution down to 1 s.
CONCLUSION:
The high sensitivity achieved increases the range of mouse SPECT applications by enabling in vivo imaging with less than a megabecquerel of tracer activity or down to 1-s frame dynamics. ...