Journal article (2022) - M. de Wit, L Gottardi, Marcel L. Ridder, K. Nagayoshi, E. Taralli, H. Akamatsu, D. Vaccaro, J. W.A. den Herder, M.P. de Bruijn, J.R. Gao
Transition-edge-sensor (TES) microcalorimeters and bolometers are used for a variety of applications.The sensors are based on the steep temperature-dependent resistance of the normal-to-superconducting transition, and are thus intrinsically sensitive to magnetic fields. Conventionally the detectors are shielded from stray magnetic fields using external magnetic shields. However, in particular for applications with strict limits on the available space and mass of an instrument, external magnetic shields might not be enough to obtain the required shielding factors or field homogeneity. Additionally, these shields are only effective for magnetic fields generated external to the TES array, and are ineffective to mitigate the impact of internally generated magnetic fields. Here we present an alternative shielding method based on a super-conducting groundplane deposited directly on the backside of the silicon nitride membrane on which the TESs are located. We demonstrate that this local shielding for external magnetic fields has a shielding factor of at the least approximately 80, and is also effective at reducing internal self-induced magnetic fields, as demonstrated by measurements and simulation of the eddy current losses in our ac-biased detectors. Measurements of 5.9-keV x-ray photons show that our shielded detectors have a high resilience to external magnetic fields, showing no degradation of the energy resolution or shifts of the energy-scale calibration for fields of several microtesla, values higher than expected in typical real-world applications. ...
Conference paper (2022) - Martin de Wit, L Gottardi, K. Nagayoshi, H. Akamatsu, M.P. de Bruijn, M.L. Ridder, E. Taralli, D. Vaccaro, J.R. Gao, More authors...
In the early 2030s, ESAs new X-ray observatory, Athena, is scheduled to be launched. It will carry two main
instruments, one of which is the X-ray Integral Field Unit (X-IFU), an X-ray imaging spectrometer, which will consist of an array of several thousand transition-edge sensors (TESs) with a proposed energy resolution of 2.5 eV for photon energies up to 7 keV. At SRON we develop the backup TES
array based on Ti/Au bilayers with a transition temperature just below 100 mK. In this contribution we will give a broad overview of the properties and capabilities of these state-of-the-art detectors. Over the years we have fabricated and studied a large number of detectors with various
geometries, providing us with a good understanding of how to precisely control the properties of our detectors. We are able to accurately vary the most important detector properties, such as the normal resistance, thermal conductance and critical temperature. This allows us to finely tune our
detectors to meet the demands of various applications. The detectors have demonstrated excellent energy resolutions of below 1.8 eV for 5.9 keV X-rays. By tuning the properties of the devices, they can be optimally matched to various read-out schemes using both AC and DC biasing. The next step is to increase the size of our TES arrays from our current kilo-pixel arrays towards the
full-sized array for X-IFU.
...
Journal article (2019) - L. Gottardi, H. Van Weers, M. L. Ridder, J. Dercksen, H. Akamatsu, M.P. de Bruijn, J. R. Gao, B. Jackson, P. Khosropanah, J. Van Der Kuur, K. Ravensberg
Dilution and adiabatic demagnetization refrigerators based on pulse tube cryocoolers are nowadays used in many low temperature physics experiments, such as atomic force and scanning tunneling microscopy, quantum computing, radiation detectors, and many others. A pulse tube refrigerator greatly simplifies the laboratory activities being a cryogen-free system. The major disadvantage of a pulse tube cooler is the high level of mechanical vibrations at the warm and cold interfaces that could substantially affect the performance of very sensitive cryogenic instruments. In this paper, we describe the performance of a very simple mechanical attenuation system used to eliminate the pulse-tube-induced low frequency noise of the superconducting transition-edge sensors under development for the instruments of the next generation of infra-red and X-ray space observatories. ...
Journal article (2019) - K. Nagayoshi, M. L. Ridder, M. P. Bruijn, L. Gottardi, E. Taralli, P. Khosropanah, H. Akamatsu, S. Visser, J. R. Gao
We are developing a transition edge sensor (TES) microcalorimeter array based on a Ti/Au superconducting bilayer, as a backup option for the X-IFU instrument on the Athena X-ray observatory. The array is read out by a frequency-division multiplexing readout system using a 1–5 MHz frequency band. Extensive research collaborations between NASA/Goddard and SRON have led to new design rules for microcalorimeters such as low resistivity of the superconductor bilayer, moderately high ohmic resistance of the TES by changing the aspect ratio and no extra metal strips. We have improved our detector fabrication process according to these design principles and produced TES arrays. Although single-pixel characterizations of these arrays are ongoing, the best energy resolution of 2.0 eV for 5.9 keV X-ray has been observed with a 120 × 20 μm2 TES with a normal resistance of 150 mΩ, biased at 2.2 MHz frequency. This shows that our Ti/Au TES array has a potential to fulfill the detector requirements of the X-IFU instrument. ...