Fabrication of a glass microfluidic gas exchanger using a femtosecond laser

More Info
expand_more

Abstract

Organ-on-chip (OoC) technology has transformed biomedical research by providing a platform to simulate physiological conditions for drug development and disease modeling. Incubator-free OoC systems offer notable advantages over traditional approaches, including enhanced adaptability and impermeability to gases. Nevertheless, achieving precise oxygen regulation remains a challenge in such systems. This thesis investigates the application of femtosecond (Fs) laser ablation to fabricate a glass microfluidic gas exchange system for accurate oxygen regulation in incubator-free OoCs.
Through successful engraving of fluidic channels and integration of an off-the-shelf optical oxygen sensor, this study highlights the efficacy of Fs laser technology in the rapid prototyping of intricate glass microfluidic devices. Despite encountering challenges such as dimensional losses and debris clogging, the study presents a functional gas exchanger prototype. Future research directions include optimization efforts, addressing issues like gas permeation through connectors, and testing under physiological conditions to further advance OoC technology.