Advancing minimally invasive surgery: A cutting-edge cable-actuated conveying mechanism for reliable tissue transportation

Journal Article (2023)
Author(s)

V.G. Kortman (TU Delft - Medical Instruments & Bio-Inspired Technology)

Y.J.M.A. Verberne (Student TU Delft)

J. Jovanova (TU Delft - Transport Engineering and Logistics)

Aimée Sakes (TU Delft - Medical Instruments & Bio-Inspired Technology)

Research Group
Medical Instruments & Bio-Inspired Technology
Copyright
© 2023 V.G. Kortman, Y.J.M.A. Verberne, J. Jovanova, A. Sakes
DOI related publication
https://doi.org/10.1371/ journal.pone.0295585
More Info
expand_more
Publication Year
2023
Language
English
Copyright
© 2023 V.G. Kortman, Y.J.M.A. Verberne, J. Jovanova, A. Sakes
Research Group
Medical Instruments & Bio-Inspired Technology
Issue number
12
Volume number
18
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

Introduction
Tissue extraction plays a crucial role in various medical disciplines, with aspiration catheters serving as the prevailing method. Unfortunately, these catheters face limitations such as clogging and dependence on tissue properties and device dimensions. Therefore, there is a pressing need for an improved tissue extraction device that enables efficient and reliable tissue removal during Minimally Invasive Surgery (MIS).

Methods
In this study, we present a novel tissue transport system that utilizes a cylindrical conveyor belt mechanism for reliable tissue transportation. We conducted experiments using a proof-of-principle prototype to explore the influence of tissue elasticity, rotational velocity, instrument orientation, and tissue shape on the transportation rate, efficiency, and reliability. Tissue phantoms with gelatine concentrations of 3, 9, and 12 wt% were employed to simulate a range of Young’s moduli from 1 to 110 kPa.

Results
The mean transportation rates for these phantoms were 7.75±0.48, 8.43±1.50, and 8.90±0.56 g/min, respectively. Notably, all phantoms were transported successfully. The perfect reliability exhibited underscores the potential of our instrument as an alternative to aspiration catheters.

Conclusion
This research presents a significant step forward in the field of tissue extraction, offering a promising approach for MIS with enhanced efficiency and reliability.