Development of a Microfluidic Sweat Rate Sensor for Continuous Monitoring in a Wearable Skin Patch

Master Thesis (2026)
Author(s)

M. Ruiz-Kaiser Ferrer (TU Delft - Mechanical Engineering)

Contributor(s)

A.Y. Rwei – Mentor (TU Delft - Applied Sciences)

M. Mastrangeli – Mentor (TU Delft - Electrical Engineering, Mathematics and Computer Science)

Faculty
Mechanical Engineering
More Info
expand_more
Publication Year
2026
Language
English
Graduation Date
30-01-2026
Awarding Institution
Delft University of Technology
Programme
Biomedical Engineering, Medical Devices
Faculty
Mechanical Engineering
Downloads counter
209
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

Wearable technologies enable non-invasive and continuous monitoring of physiological parameters offering cost-efficient alternatives to clinical assessments. Among them, sweat sensors expand the available biometrics by providing access to biochemical information. Analyte concentrations in sweat vary substantially with inter- and intra individual differences in sweat rate (SR). Therefore, sweat sensors need to measure SR to deliver reliable data. This work presents the design and development of a microfluidic SR sensor for continuous monitoring in a skin patch. The device uses electrodes patterned on polyimide and embedded within polydimethylsiloxane (PDMS) microfluidic channels to quantify SR by tracking fluid progression through impedance measurements. Two electrode materials, thin-film gold electrodes and laser-induced graphene (LIG) electrodes, were investigated for this application. Gold electrodes showed a more stable SR sensing performance demonstrating high sensitivity across physiologically relevant SRs, ranging from values typical at rest (0.2 µL/min) to physical exercise (5 µL/min). Additionally, the sensor maintained a stable performance under mechanical deformation as well as during on-body testing. This supports the sensor’s suitability for wearable applications. The sensor response was also evaluated across electrolyte concentrations (10-100 mM) and temperature variations (20-40 ºC), highlighting the need to include calibration strategies that can compensate for these effects. The proposed sensor architecture was designed to be compatible alongside ion-selective electrodes, for sodium and chloride sensing, to support future incorporation into a complete wearable sweat analysis platform.

Files

License info not available
warning

File under embargo until 30-01-2027