The influence of prolonged instrument manipulation on gas leakage through trocars

Journal Article (2023)
Author(s)

Daniel Robertson (TU Delft - Medical Instruments & Bio-Inspired Technology)

Matthijs van Duijn (Student TU Delft)

Alberto Arezzo (University of Turin)

Yoav Mintz (Hadassah-Hebrew University Medical Center)

T Horeman (TU Delft - Medical Instruments & Bio-Inspired Technology)

G.B. More authors (External organisation)

Research Group
Medical Instruments & Bio-Inspired Technology
Copyright
© 2023 P.D. Robertson, Matthijs van Duijn, Alberto Arezzo, Yoav Mintz, T. Horeman, More Authors
DOI related publication
https://doi.org/10.1007/s00464-023-10240-5
More Info
expand_more
Publication Year
2023
Language
English
Copyright
© 2023 P.D. Robertson, Matthijs van Duijn, Alberto Arezzo, Yoav Mintz, T. Horeman, More Authors
Research Group
Medical Instruments & Bio-Inspired Technology
Issue number
9
Volume number
37
Pages (from-to)
7325-7335
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

Background: During laparoscopic surgery, CO2 insufflation gas could leak from the intra-abdominal cavity into the operating theater. Medical staff could therefore be exposed to hazardous substances present in leaked gas. Although previous studies have shown that leakage through trocars is a contributing factor, trocar performance over longer periods remains unclear. This study investigates the influence of prolonged instrument manipulation on gas leakage through trocars. Methods: Twenty-five trocars with diameters ranging from 10 to 15 mm were included in the study. An experimental model was developed to facilitate instrument manipulation in a trocar under loading. The trocar was mounted to a custom airtight container insufflated with CO2 to a pressure of 15 mmHg, similar to clinical practice. A linear stage was used for prolonged instrument manipulation. At the same time, a fixed load was applied radially to the trocar cannula to mimic the reaction force of the abdominal wall. Gas leakage was measured before, after, and during instrument manipulation. Results: After instrument manipulation, leakage rates per trocar varied between 0.0 and 5.58 L/min. No large differences were found between leakage rates before and after prolonged manipulation in static and dynamic measurements. However, the prolonged instrument manipulation did cause visible damage to two trocars and revealed unintended leakage pathways in others that can be related to production flaws. Conclusion: Prolonged instrument manipulation did not increase gas leakage rates through trocars, despite damage to some individual trocars. Nevertheless, gas leakage through trocars occurs and is caused by different trocar-specific mechanisms and design issues.