Laparoscopic versus robot-assisted left hemicolectomy
A pilot study on sustainability
Background: The National Health System is responsible for 8–10% of total greenhouse gas emissions. Operating rooms are responsible for 60–70% of all hospital waste. Over the last 30 years abdominal surgery transcended from a laparoscopic approach toward a robot-assisted approach. The role of robot-assisted laparoscopic surgery is still debated in some procedures, such as colorectal surgery. The studies available in scientific literature comparing laparoscopic and robot-assisted left hemicolectomy are focused on clinical outcomes. The environmental sustainability of these procedures remains largely unexplored, representing a key area that our study seeks to investigate. Methods: In this pilot study consecutive patients scheduled for a minimally invasive left hemicolectomy for diverticular disease or cancer were recruited and randomly assigned 1:1 to the laparoscopic or robotic groups. The “Green Team” supported the operating room staff in separate waste collection during the surgical procedures. Primary end point was CO2 consumption and secondary endpoints the specific mass of the most important waste stream. Results: Ten patients were enrolled. Robot-assisted left hemicolectomy required more CO2 consumption in liters to maintain pneumoperitoneum (p = 0.03) compared with laparoscopic left hemicolectomy and required a longer operation time (p = 0.04). In total, the robot and laparoscopic approaches produced a total of 74.5 and 54 kg of plastic, non-woven fabric (TNT), unsorted waste bins, and biohazardous waste combined, which cost €92 and €71 to dispose of. Conclusion: Robot-assisted left hemicolectomy seems to have a greater environmental impact compared with laparoscopic left hemicolectomy in terms of both CO2 emissions and waste production. Given the growing focus on operating room sustainability, further studies are needed to compare laparoscopic and robotic techniques to inform surgical decisions.
“O.R. GOES GREEN”
A first step toward reducing our carbon footprint in the operating room and hospital
Hospitals in Europe produce approximately 6 million tons of medical waste annually, about one-third of this originating in operating rooms. Most of it is solid waste, which can be recycled if bodily fluids do not contaminate it. Only 2–3% of hospital waste must be disposed of as infectious waste, and this is much lower than the 50–70% of garbage in the biohazard waste stream. In June 2021, at the main operating room of the Department of General Surgery of the University of Turin, we began a separate collection program for materials consisting of plastic, paper, TNT (material not contaminated by bodily fluids), and biohazardous waste. We calculated the number of boxes and the weight of special waste disposed produced every month in one operating room for 18 months. The monthly number of Sanibox and the monthly weight of biohazardous waste decreased during the observation period. The reduction trend was not constant but showed variations during the 18 months. Direct proportionality between number of low-complexity procedures and production of biohazardous waste was found (p = 0.050). We observed an optimization in the collection and filling of plastic, paper and TNT boxes separated and sent for recycling. One of the barriers to recycling hospital waste, and surgical waste in particular, is the failure to separate infectious waste from clean waste. A careful separate collection of waste in the operating room is the first step in reducing environmental pollution and management costs for the disposal of hospital waste.
Surgeons performing robotic-assisted laparoscopic surgery experience physical stress and overuse of shoulder muscles due to sub-optimal arm support during surgery. The objective is to present a novel design and prototype of a dynamic arm support for robotic laparoscopic surgery to evaluate its ergonomics and performance on the AdLap-VR simulation training device. The prototype was designed using the mechanical engineering design process: Technical requirements, concept creation, concept selection, 3D-design and built of the prototype. A crossover study was performed on a marble sorting task on the AdLap-VR. The first group performed four trials without the arm support, followed by four trials with the arm support, and the other group executed the sequence vice versa. The performance parameters used were time to complete (s), path length (mm), and the number of collisions. Afterward, the participants filled out a questionnaire on the ergonomic experience regarding both situations. 20 students executed 160 performed trials on the AdLap-VR Significant decreases in the subjective comfort parameters mental demand, physical demand, effort and frustration were observed as a result of introducing the novel arm support. Significant decreases in the objective performance parameters path length and the number of collisions were also observed during the tests. The newly developed dynamic arm support was found to improve comfort and enhance performance through increased stability on the robotic surgery skills simulator AdLap-VR.
Background: During minimal access surgery, surgical smoke is produced which can potentially be inhaled by the surgical team, leading to several health risks. This smoke can escape from the abdominal cavity into the operating room due to trocar leakage. The trocars and insufflator that are used during surgery influence gas leakage. Therefore, this study compares particle escape from a valveless (Conmed AirSeal iFS), and a conventional (Karl Storz Endoflator) system. Materials and methods: Using an in vitro model, a conventional and a valveless trocar system were compared. A protocol that simulated various surgical phases was defined to assess the surgical conditions and particle leakage. Insufflation pressures and instrument diameters were varied as these are known to affect gas leakage. Results: The conventional trocar leaked during two distinct phases. Removal of the obturator caused a sudden release of particles. During instrument insertion, an average of 211 (IQR 111) particles per second escaped when using the 5 mm diameter instrument. With the 10 mm instrument, 50 (IQR 13) particles per second were measured. With the conventional trocar, a higher abdominal pressure increased particle leakage. The valveless trocar demonstrated a continuously high particle release during all phases. After the obturator was removed, particle escape increased sharply. Particle escape decreased to 1276 (IQR 580) particles per second for the 5 mm instrument insertion, and 1084 (IQR 630) particles per second for 10 mm instrument insertion. With the valveless trocar system, a higher insufflation pressure lowered particle escape. Conclusions: This study shows that a valveless trocar system releases more particles into the operating room environment than a conventional trocar. During instrument insertion, the leakage through the valveless system is 6 to 20 times higher than the conventional system. With a valveless trocar, leakage decreases with increasing pressure. With both trocar types leakage depends on instrument diameter.
Background: During laparoscopic surgery, CO2 insufflation gas could leak from the intra-abdominal cavity into the operating theater. Medical staff could therefore be exposed to hazardous substances present in leaked gas. Although previous studies have shown that leakage through trocars is a contributing factor, trocar performance over longer periods remains unclear. This study investigates the influence of prolonged instrument manipulation on gas leakage through trocars. Methods: Twenty-five trocars with diameters ranging from 10 to 15 mm were included in the study. An experimental model was developed to facilitate instrument manipulation in a trocar under loading. The trocar was mounted to a custom airtight container insufflated with CO2 to a pressure of 15 mmHg, similar to clinical practice. A linear stage was used for prolonged instrument manipulation. At the same time, a fixed load was applied radially to the trocar cannula to mimic the reaction force of the abdominal wall. Gas leakage was measured before, after, and during instrument manipulation. Results: After instrument manipulation, leakage rates per trocar varied between 0.0 and 5.58 L/min. No large differences were found between leakage rates before and after prolonged manipulation in static and dynamic measurements. However, the prolonged instrument manipulation did cause visible damage to two trocars and revealed unintended leakage pathways in others that can be related to production flaws. Conclusion: Prolonged instrument manipulation did not increase gas leakage rates through trocars, despite damage to some individual trocars. Nevertheless, gas leakage through trocars occurs and is caused by different trocar-specific mechanisms and design issues.
Safe implementation of hand held steerable laparoscopic instruments
A survey among EAES surgeons
The complexity of handheld steerable laparoscopic instruments (SLI) may impair the learning curve compared to conventional instruments when first utilized. This study aimed to provide the current state of interest in the use of SLI, the current use of these in daily practice and the type of training which is conducted before using SLI in the operating room (OR) on real patients. An online survey was distributed by European Association of Endoscopic Surgery (EAES) Executive Office to all active members, between January 4th and February 3rd, 2020. The survey consisted of 14 questions regarding the usage and training of steerable laparoscopic instruments. A total of 83 members responded, coming from 33 different countries. Twenty three percent of the respondents using SLI, were using the instruments routinely and of these 21% had not received any formal training in advance of using the instruments in real patients. Of all responding EAES members, 41% considered the instruments to potentially compromise patient safety due to their complexity, learning curve and the inexperience of the surgeons. The respondents reported the three most important aspects of a possible steerable laparoscopic instruments training curriculum to be: hands-on training, safe tissue handling and suturing practice. Finally, a major part of the respondents consider force/pressure feedback data to be of significant importance for implementation of training and assessment of safe laparoscopic and robotic surgery. Training and assessment of skills regarding safe implementation of steerable laparoscopic instruments is lacking. The respondents stressed the need for specific hands-on training during which feedback and assessment of skills should be guaranteed before operating on real patients.
During laparoscopy, the abdominal cavity is insufflated with carbon dioxide (CO2) that could become contaminated with viruses and surgical smoke. Medical staff is potentially exposed when this gas leaks into the operating room through the instruments and past trocar valves. No detailed studies currently exist that have quantified these leakage pathways. Therefore, the goal of this study was to quantify the gas leakages through trocars and instruments, during minimally invasive procedures.
Methods
A model of the surgical environment was created, consisting of a rigid container with an interface for airtight clamping of laparoscopic equipment such as trocars and surgical instruments. The model was insufflated to 15 mm Hg using a pressure generator and a pneumotachograph measured the equipment gas leak. A protocol of several use cases was designed to simulate the motions and forces the surgeon exerts on the trocar during surgery.
Results
Twenty-three individual trocars and twenty-six laparoscopic instruments were measured for leakage under the different conditions of the protocol. Trocar leakages varied between 0 L/min and more than 30 L/min, the instruments revealed a range of leakages between 0 L/min and 5.5 L/min. The results showed that leakage performance varied widely between trocars and instruments and that the performance and location of the valves influenced trocar leakage.
Conclusions
We propose trocar redesigns to overcome specific causes of gas leaks. Moreover, an international testing standard for CO2 leakage for all new trocars and instruments is needed so surgical teams can avoid this potential health hazard when selecting new equipment. ...
During laparoscopy, the abdominal cavity is insufflated with carbon dioxide (CO2) that could become contaminated with viruses and surgical smoke. Medical staff is potentially exposed when this gas leaks into the operating room through the instruments and past trocar valves. No detailed studies currently exist that have quantified these leakage pathways. Therefore, the goal of this study was to quantify the gas leakages through trocars and instruments, during minimally invasive procedures.
Methods
A model of the surgical environment was created, consisting of a rigid container with an interface for airtight clamping of laparoscopic equipment such as trocars and surgical instruments. The model was insufflated to 15 mm Hg using a pressure generator and a pneumotachograph measured the equipment gas leak. A protocol of several use cases was designed to simulate the motions and forces the surgeon exerts on the trocar during surgery.
Results
Twenty-three individual trocars and twenty-six laparoscopic instruments were measured for leakage under the different conditions of the protocol. Trocar leakages varied between 0 L/min and more than 30 L/min, the instruments revealed a range of leakages between 0 L/min and 5.5 L/min. The results showed that leakage performance varied widely between trocars and instruments and that the performance and location of the valves influenced trocar leakage.
Conclusions
We propose trocar redesigns to overcome specific causes of gas leaks. Moreover, an international testing standard for CO2 leakage for all new trocars and instruments is needed so surgical teams can avoid this potential health hazard when selecting new equipment.