X-ray photon-counting using silicon photomultiplier-based scintillation detectors at high x-ray tube currents
S.J. Van Der Sar (TU Delft - RST/Medical Physics & Technology)
Stefan E. Brunner (Broadcom Inc., Regensburg)
Dennis Schaart (TU Delft - RST/Medical Physics & Technology, HollandPTC)
More Info
expand_more
Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.
Abstract
We investigate X-ray photon-counting scintillation detectors with silicon photomultiplier (SiPM) readout. These circumvent some drawbacks of direct-conversion detectors. We measured observed count rate (OCR) versus X-ray tube current for single-pixel detectors consisting of LYSO:Ce and YAP:Ce scintillators coupled to ultrafast SiPMs. For a 30 keV threshold, the maximum OCRs equal 4.5 Mcps/pixel (LYSO:Ce) and 5.5 Mcps/pixel (YAP:Ce) for paralyzable-like counting and 10 Mcps/pixel (LYSO:Ce) and 12.5 Mcps/pixel (YAP:Ce) for nonparalyzable-like counting. We estimate that the twice as fast LaBr3:Ce scintillator yields OCRs approaching those of CdTe/CZT-based photon-counting CT detectors. We also show energy response data and discuss dose-efficient pixel miniaturization.