Recovering the Acoustic Change Complex: A Single-Trial Pipeline for Reduced EEG Recording Time

Master Thesis (2026)
Author(s)

G. Khosrokhavar (TU Delft - Electrical Engineering, Mathematics and Computer Science)

Contributor(s)

R.C. Hendriks – Mentor (TU Delft - Electrical Engineering, Mathematics and Computer Science)

Jantien Vroegop – Mentor (Erasmus MC)

A. Savva – Graduation committee member (TU Delft - Electrical Engineering, Mathematics and Computer Science)

Faculty
Electrical Engineering, Mathematics and Computer Science
More Info
expand_more
Publication Year
2026
Language
English
Graduation Date
03-07-2026
Awarding Institution
Delft University of Technology
Programme
Electrical Engineering, Signals and Systems
Faculty
Electrical Engineering, Mathematics and Computer Science
Page Views
131
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

The Acoustic Change Complex (ACC) is a cortical auditory Event-Related Potential (ERP) used as an objective measure of auditory discrimination. Its clinical adoption is limited by the long recording times required to recover the ACC waveform from noisy single-trial EEG through conventional trial averaging.

This thesis develops a two-part single-trial analysis pipeline. Part 1 classifies each trial as Clear ACC, Possible ACC, or Noise Dominant using rule-based hard-gate filtering followed by template-based scoring. Part 2 corrects per-trial latency offsets using four alignment algorithms: direct N1-peak alignment, direct P2-peak alignment, the Woody filter, and dynamic time warping (DTW).

Evaluated on seven normal-hearing adults at electrode E6, the pipeline raises the averaged N1–P2 amplitude from 2.8 µV (grand average of all trials) to 9.9 µV for the best alignment method, a 3.5× improvement. The Woody filter yields the highest absolute amplitude, while direct P2-peak alignment yields the highest noise-corrected amplitude. Validation against non-hearing acquisitions yields a mean specificity of 96.4%, confirming that the detection pipeline does not produce substantial spurious detections on stimulus-absent EEG.

This work contributes toward a more efficient ACC recording framework by showing that trial-level information can be used to enhance waveform recovery and reduce dependence on long averaging protocols.

Files

License info not available
warning

File under embargo until 01-08-2027