A 1.55 μW Bio-Impedance Measurement System for Implantable Cardiac Pacemakers in 0.18 μm CMOS

Journal Article (2018)
Author(s)

M. Zamani (University of Tehran)

Y. Rezaeiyan (University of Tehran)

Omid Shoaei (University of Tehran)

Wouter A. Serdijn (TU Delft - Bio-Electronics)

Research Group
Bio-Electronics
DOI related publication
https://doi.org/10.1109/TBCAS.2017.2776528
More Info
expand_more
Publication Year
2018
Language
English
Research Group
Bio-Electronics
Issue number
1
Volume number
12
Pages (from-to)
211-221

Abstract

This paper presents an implantable bio-impedance measurement system for cardiac pacemakers. The fully integrated system features a lowpower analog front-end and pulse widthmodulated output. The bio-impedance readout benefits from voltage to time conversion to achieve a very low power consumption for wirelessly transmitting the data outside the body. The proposed IC is fabricated in a 0.18μmCMOSprocess and is capable ofmeasuring the bio-impedance at 2 kHz over a wide dynamic range from 100λ to 3.3 kλ with 1.35 λrms accuracy and 1 μA maximum current injection while consuming just 1.55 μA from a 1 V supply.