A 1.55 μW Bio-Impedance Measurement System for Implantable Cardiac Pacemakers in 0.18 μm CMOS
M. Zamani (University of Tehran)
Y. Rezaeiyan (University of Tehran)
Omid Shoaei (University of Tehran)
Wouter A. Serdijn (TU Delft - Bio-Electronics)
More Info
expand_more
Abstract
This paper presents an implantable bio-impedance measurement system for cardiac pacemakers. The fully integrated system features a lowpower analog front-end and pulse widthmodulated output. The bio-impedance readout benefits from voltage to time conversion to achieve a very low power consumption for wirelessly transmitting the data outside the body. The proposed IC is fabricated in a 0.18μmCMOSprocess and is capable ofmeasuring the bio-impedance at 2 kHz over a wide dynamic range from 100λ to 3.3 kλ with 1.35 λrms accuracy and 1 μA maximum current injection while consuming just 1.55 μA from a 1 V supply.