A simple and effective attachment to prevent dome diffraction reaching 2D or 1D detectors in x-ray diffractometers
N.M. van der Pers (TU Delft - Team Amarante Bottger)
R. W.A. Hendrikx (TU Delft - Team Amarante Bottger)
Amarante J. Bottger (TU Delft - Team Amarante Bottger)
More Info
expand_more
Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.
Abstract
An attachment has been developed for x-ray diffractometer systems equipped with a domed stage when using a 2D or 1D detector. It consists of a single screen in front of the detector positioned such that it blocks diffraction from the dome. This results in measured data free of disturbing spurious peaks and background, thereby greatly facilitating further data analysis. Its working principle is universally applicable and allows for all specimen orientation movements needed for x-ray diffraction measurements, including texture, stress, and mapping.