Journal article (2026) - Raphael Panskus, Andrada Iulia Velea, Lukas Holzapfel, Christos Pavlou, Flora Nelissen, Rick Waasdorp, David Maresca, Valeria Gazzola, Vasiliki Giagka, More authors...
Neural interfaces that unify diagnostic and therapeutic functionalities hold particular promise for advancing both fundamental neuroscience and clinical neurotechnology. Functional ultrasound imaging (fUSI) has recently emerged as a powerful modality for high-resolution, non-invasive monitoring of brain function and structure. However, conventional metal-based microelectrodes typically impede ultrasound propagation, limiting compatibility with fUSI. Here, we present flexible, ultrasound-transparent neural interfaces that retain practical metal thicknesses while achieving high acoustic transparency. We introduce a theoretical and simulation-based framework to investigate the conditions under which commonly used polymers and metals in neural interfaces can become acoustically transparent. Based on these insights, we propose design guidelines that maximise ultrasound transmission through soft neural interfaces. We experimentally validate our approach through immersion experiments and by demonstrating the acoustic transparency of a suitably engineered interface using fUSI in phantom and in vivo experiments. Finally, we discuss the potential extension of this approach to therapeutic focused ultrasound (FUS). This work establishes a foundation for the development of multimodal neural interfaces with enhanced diagnostic and therapeutic capabilities, enabling both scientific discovery and translational impact. ...
Journal article (2025) - Andrada I. Velea, Raphael Panskus, Benedikt Szabo, Vera A.L. Oppelt, Lukas Holzapfel, Cyril B. Karuthedath, Abhilash T. Sebastian, Thomas Stieglitz, Alessandro S. Savoia, Vasiliki Giagka
Ultrasound (US) is a promising modality for wirelessly powering implantable devices, requiring encapsulated receivers to ensure long-term stability. Traditional hermetic packaging often limits acoustic transmission, making polymer-based encapsulation a more suitable alternative. This study investigates how implant-grade polymers, thermoplastic polyurethane (TPU), parylene-C, and medical-grade silicones (MED-1000 and MED2-4213), affect the receive performance of piezoelectric micromachined ultrasonic transducers (PMUTs). Simulations and measurements between 1 and 7 MHz show that all tested materials exhibit transmission coefficients above 94% at nanometer- and micrometer-scale thicknesses, confirming their acoustic transparency. The results show that although coated PMUTs are acoustically well matched with the surrounding water medium, the added mechanical load of the coating can hinder membrane motion and reduce the energy transferred to the PMUTs. Modeling and experimental data demonstrate that stiffer coatings, such as parylene-C, lead to a reduced sensitivity when similar thicknesses are used. Likewise, residual stress in materials like MED-1000 can also degrade the performance. These effects are not evident from acoustic transmission measurements alone, underscoring the need to assess both acoustic and mechanical properties when selecting encapsulation materials. In general, softer materials offer excellent acoustic performance for PMUT encapsulation, while stiffer materials must be applied in thinner layers to avoid impairing PMUT function. ...
Journal article (2024) - Lukas Holzapfel, Vasiliki Giagka
Traditionally, implants are powered by batteries, which have to be recharged by an inductive power link. In the recent years, ultrasonic power links are being investigated, promising more available power for deeply implanted miniaturized devices. These implants often need to transfer back information. For ultrasonically powered implants, this is usually achieved with on-off keying (OOK) based on backscatter modulation, or active driving of a secondary transducer. In this article, we propose to superimpose subcarriers, effectively leveraging frequency-shift keying (FSK), which increases the robustness of the link against interference and fading. It also allows for simultaneous powering and communication, and inherently provides the possibility of frequency domain multiplexing for implant networks. The modulation scheme can be implemented in miniaturized application-specific integrated circuits (ASICs), field programmable gate arrays (FPGAs), and microcontrollers. We have validated this modulation scheme in a water tank during continuous ultrasound and movement. We achieved symbol rates of up to 104 kBd, and were able to transfer data through 20 cm of water and through a 5 cm tissue phantom with additional misalignment and during movements. This approach could provide a robust uplink for miniaturized implants that are located deep inside the body and need continuous ultrasonic powering. ...
Conference paper (2023) - Jana M. Späth, Konstantina Kolovou Kouri, Lukas Holzapfel, Roland Thewes, Vasiliki Giagka
Safety is a critical consideration when designing an electrical neural stimulator, given the direct contact with neural tissue. This paper presents the design of a charge balancing system suitable for frequencies up to the kilohertz domain, to be used as an add-on system for stimulators over a wide range of frequencies, also covering nerve conduction blocking. It operates independently of the stimulator timing by continuously sensing the offset voltage, and applying a corrective current to the electrode, using the offset compensation technique. To ensure its stand-alone capability, the system is battery-powered, and includes a safety and start-up circuit. Electrical measurements verified the functionality of the circuit, demonstrating a residual offset of only 0.7 mV for 1 V biphasic pulses at 50 kHz. When tested for 20 kHz biphasic pulse at a 5 V amplitude, the offset was measured at -11.6 mV, which is still within the (commonly used) ±50 mV safety window. ...

An Experimental Comparison Between Binary-Frequency-Shift-Keying and On-Off-Keying with Backscatter Modulation

Conference paper (2023) - Lukas Holzapfel, Vasiliki Giagka
Implantable devices need to communicate information to the outside world. For deep-seated miniaturized implants ultrasound communication can be favourable. However, implants need to operate during movement, and the selected communication scheme should be assessed accordingly. In this work, we implemented a simple protocol to transfer data packets based on On-Off-Keying (OOK) and Frequency-Shift-Keying (FSK) by backscatter modulation in ultrasonic communication links for deep implants.We then used it to compare FSK vs OOK encoding regarding the bit error rate during continuous ultrasound power transfer, and while moving, in a water tank setup. Our experiment shows, that sub-millimeter movements can have severe effects on OOK communication, but not for FKS. Therefore, FSK can allow for backscatter communication from deep implants regardless of their position and involved movements. The protocol can also be adapted to other backscatter modulation schemes in the future. ...
Conference paper (2023) - Raphael Panskus, Lukas Holzapfel, Wouter A. Serdijn, Vasiliki Giagka
Recording neuronal activity triggered by electrical impulses is a powerful tool in neuroscience research and neural engineering. It is often applied in acute electrophysiological experimental settings to record compound nerve action potentials. However, the elicited neural response is often distorted by electrical stimulus artifacts, complicating subsequent analysis. In this work, we present a model to better understand the effect of the selected amplifier configuration and the location of the ground electrode in a practical electrophysiological nerve setup. Simulation results show that the stimulus artifact can be reduced by more than an order of magnitude if the placement of the ground electrode, its impedance, and the amplifier configuration are optimized. We experimentally demonstrate the effects in three different settings, in-vivo and in-vitro. ...