The design of feasible trajectories to traverse the k-space for sampling in magnetic resonance imaging (MRI) is important while considering ways to reduce the scan time. Over the recent years, non-Cartesian trajectories have been observed to result in benign artifacts and being l ...
Variable density sampling of the k-space in MRI is an integral part of trajectory design. It has been observed that data-driven trajectory design methods provide a better image reconstruction as compared to trajectories obtained from a fixed or a parametric density function. In t ...
A novel class of k-space trajectories for magnetic resonance imaging (MRI) sampling using space filling curves (SFCs) is presented here. More specifically, Peano, Hilbert and Sierpinski curves are used. We propose 1-shot and 4-shot variable density SFCs by utilizing the space cov ...
The development of compressed sensing (CS) techniques for magnetic resonance imaging (MRI) is enabling a speedup of MRI scanning. To increase the incoherence in the sampling, a random selection of points on the k-space is deployed and a continuous trajectory is obtained by solvin ...

Contributed

1 records found

Electromagnetic Fields in MRI

Analytical Methods and Applications

Electrical properties, the conductivity and permittivity of tissue, are quantities that describe the interaction of an object and electromagnetic fields. These properties influence electromagnetic fields and are influenced themselves by physiological phe- nomena such as lesions o ...