Myocardial Arterial Spin Labeling with Double Inversion Recovery for reduced physiological noise

Journal Article (2025)
Author(s)

Maša Božić-Iven (University of Heidelberg, TU Delft - ImPhys/Weingärtner group)

Stanislas Rapacchi (University Hospital of Vaud (CHUV))

Yi Zhang (TU Delft - ImPhys/Tao group)

Qian Tao (TU Delft - ImPhys/Tao group)

Lothar Rudi Schad (University of Heidelberg)

Sebastian Weingärtner (TU Delft - ImPhys/Computational Imaging, TU Delft - ImPhys/Weingärtner group)

Research Group
ImPhys/Tao group
DOI related publication
https://doi.org/10.1002/mrm.70018
More Info
expand_more
Publication Year
2025
Language
English
Research Group
ImPhys/Tao group
Issue number
6
Volume number
94
Pages (from-to)
2460-2474
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

Purpose: To introduce Double Inversion Recovery (DIR) preparations for myocardial Arterial Spin Labeling (myoASL) for mitigation of heart rate (HR) variability induced physiological noise (PN).

Methods: DIR-labeling was implemented for double ECG-gated myoASLsequences and compared with conventional Flow-sensitive Alternating Inversion Recovery (FAIR) labeling using single inversions. In DIR-preparations, the FAIR-inversion pulses were immediately followed by an identical reinversion pulse, applied either slice-selectively or nonselectively. Bloch-equation-based simulation and phantom experiments were performed to evaluate the PN and SNR across a range of HR variabilities. Data from six healthy subjects were acquired to evaluate myocardial blood flow (MBF), PN, and SNR in vivo.

Results: Simulation experiments showed that the averageMBFvalues remained nearly constant across the range of HR variabilities and were comparable across all three sequences. However, DIR-labeling allowed for greater recovery of the myocardial background signal, which mitigates the sensitivity to HR-dependent changes in the inversion time. Consequently, PN in the presence of HR variability was substantially reduced with DIR-labeling. For HR variabilities corresponding to the mean value observed in vivo, this resulted in a simulated SNR gain of 1.79 ± 0.90 for selective and 1.55 ± 0.77 for nonselective DIR-labeling. In vivo, DIR-labeling showed reduced PN, with 53% (p < 0.05)/44% (p = 0.16) less PN compared with conventional FAIR-myoASL, leading to an average SNR gain of 1.47 ± 0.63 (p = 0.09)/1.32 ± 0.57 (p = 0.84) with selective/nonselective reinversions.

Conclusion: The proposed DIR-preparations reduce sensitivity to HR variations and alleviate PN in double ECG-gated myoASL, improving the precision of myoASL-based perfusion quantification.