Feasibility Study for a High-Frequency Flexible Ultrasonic Cuff for High-Precision Vagus Nerve Ultrasound Neuromodulation

Journal Article (2024)
Authors

Cornelis Van Damme (Student TU Delft)

Gandhika K. Wardhana (TU Delft - Bio-Electronics)

A. Velea (TU Delft - Bio-Electronics, Fraunhofer Institute for Reliability and Microintegration IZM)

Vasiliki Giagka (Fraunhofer Institute for Reliability and Microintegration IZM, TU Delft - Bio-Electronics)

T.M. Costa (TU Delft - Bio-Electronics)

Research Group
Bio-Electronics
To reference this document use:
https://doi.org/10.1109/TUFFC.2024.3381923
More Info
expand_more
Publication Year
2024
Language
English
Research Group
Bio-Electronics
Issue number
7
Volume number
71
Pages (from-to)
745-756
DOI:
https://doi.org/10.1109/TUFFC.2024.3381923
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

In the emerging research field of bioelectronic medicine, it has been indicated that neuromodulation of the vagus nerve (VN) has the potential to treat various conditions such as epilepsy, depression, and autoimmune diseases. In order to reduce side effects, as well as to increase the effectiveness of the delivered therapy, sub-fascicle stimulation specificity is required. In the electrical domain, increasing spatial selectivity can only be achieved using invasive and potentially damaging approaches like compressive forces or nerve penetration. To avoid these invasive methods while obtaining a high spatial selectivity, a 2-mm diameter extraneural cuff-shaped proof-of-concept design with integrated lead zirconate titanate (PZT) based ultrasound (US) transducers is proposed in this article. For the development of the proposed concept, wafer-level microfabrication techniques are employed. Moreover, acoustic measurements are performed on the device, in order to characterize the ultrasonic beam profiles of the integrated PZT-based US transducers. A focal spot size of around 200× 200 μ m is measured for the proposed cuff. Moreover, the curvature of the device leads to constructive interference of the US waves originating from multiple PZT-based US transducers, which in turn leads to an increase of 45% in focal pressure compared to the focal pressure of a single PZT-based US transducer. Integrating PZT-based US transducers in an extraneural cuff-shaped design has the potential to achieve high-precision US neuromodulation of the VN without requiring intraneural implantation.

Files

Feasibility_Study_for_a_High-F... (pdf)
(pdf | 1.93 Mb)
- Embargo expired in 30-09-2024
License info not available