KR
K. Rider
4 records found
1
ICON Far UltraViolet (FUV) imager contributes to the ICON science objectives by providing remote sensing measurements of the daytime and nighttime atmosphere/ionosphere. During sunlit atmospheric conditions, ICON FUV images the limb altitude profile in the shortwave (SW) band at
...
V-UV spectrographic imager (FUV) for Icon mission
From optical design to vacuum calibration
The ICON mission is led by the University of California-Berkeley (Space Sciences Laboratory). In the frame of this mission the Space Center of Liege was involved in the optical design optimization and related analysis, and VUV on ground calibration.
Alignment and calibration of the ICON-FUV instrument
Development of a vacuum UV facility
The optical calibration of the ICON-FUV instrument requires designing specific ground support equipment (GSE). The ICON-FUV instrument is a spectrographic imager that operates on two specific wavelengths in the UV (135.6 nm and 157 nm). All the operations have to be performed und
...
ICON FUV is a two channel spectrographic imager that measures intensity and spatial distribution of oxygen (135.6 nm) and molecular nitrogen (157 nm) of the ionosphere. As those wavelengths are strongly absorbed by the atmosphere, the optical elements of the system have to be tes
...