A 0.91mW/Element Pitch-Matched Front-End ASIC with Integrated Subarray Beamforming ADC for Miniature 3D Ultrasound Probes
C. Chen (TU Delft - Electronic Instrumentation)
Z. Chen (TU Delft - Electronic Instrumentation)
D. Bera (Erasmus MC)
E.C. Noothout (ImPhys/Acoustical Wavefield Imaging )
Z.Y. Chang (TU Delft - Electronic Instrumentation)
M. Tan (TU Delft - Electronic Instrumentation)
Hendrik J. Vos (Erasmus MC)
J. G. Bosch (Erasmus MC)
M.D. Verweij (Erasmus MC, ImPhys/Acoustical Wavefield Imaging )
N. de Jong (Erasmus MC, ImPhys/Acoustical Wavefield Imaging )
M.A.P. Pertijs (TU Delft - Electronic Instrumentation)
More Info
expand_more
Abstract
Data acquisition from 2D transducer arrays is one of the main challenges for the development of emerging miniature 3D ultrasound imaging devices, such as 3D trans-esophageal (TEE) and intra-cardiac echocardiography (ICE) probes (Fig. 10.5.1). The main obstacle lies in the mismatch between the large number of transducer elements (103 to 104) and the limited cable count (<;200). Recent advances in transducer-on-CMOS integration have enabled the use of in-probe subarray beamforming based on delay-and-sum (DAS) circuits [1] to reduce the channel count by an order of magnitude. Further reduction calls for in-probe digitization to enable more advanced data processing and compression in the digital domain. However, prior designs [2-4] compromise on transducer pitch (> half wavelength) to accommodate the ADC and consume >9mW/element, which translates into unacceptable self-heating in miniature 3D probes.
No files available
Metadata only record. There are no files for this record.