A Pitch-Matched Transceiver ASIC With Shared Hybrid Beamforming ADC for High-Frame-Rate 3-D Intracardiac Echocardiography

Journal Article (2022)
Author(s)

Y. Hopf (TU Delft - Electronic Instrumentation)

Boudewine W. Ossenkoppele (Thoraxcenter)

Mehdi Soozande (Thoraxcenter)

E. Noothout (TU Delft - ImPhys/Medical Imaging)

Z.Y. Chang (TU Delft - Electronic Instrumentation)

Chao Chen (Thoraxcenter)

H.J. Vos (TU Delft - ImPhys/Medical Imaging)

J. G. Bosch (Thoraxcenter)

M. Verweij (TU Delft - ImPhys/Medical Imaging)

N. de Jong (TU Delft - ImPhys/Medical Imaging)

M. Pertijs (TU Delft - Electronic Instrumentation)

Research Group
Electronic Instrumentation
Copyright
© 2022 Y.M. Hopf, Boudewine W. Ossenkoppele, Mehdi Soozande, E.C. Noothout, Z.Y. Chang, Chao Chen, H.J. Vos, Johan G. Bosch, M.D. Verweij, N. de Jong, M.A.P. Pertijs
DOI related publication
https://doi.org/10.1109/JSSC.2022.3201758
More Info
expand_more
Publication Year
2022
Language
English
Copyright
© 2022 Y.M. Hopf, Boudewine W. Ossenkoppele, Mehdi Soozande, E.C. Noothout, Z.Y. Chang, Chao Chen, H.J. Vos, Johan G. Bosch, M.D. Verweij, N. de Jong, M.A.P. Pertijs
Research Group
Electronic Instrumentation
Issue number
11
Volume number
57
Pages (from-to)
3228-3242
Reuse Rights

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.

Abstract

In this article, an application-specific integrated circuit (ASIC) for 3-D, high-frame-rate ultrasound imaging probes is presented. The design is the first to combine element-level, high-voltage (HV) transmitters and analog front-ends, subarray beamforming, and in-probe digitization in a scalable fashion for catheter-based probes. The integration challenge is met by a hybrid analog-to-digital converter (ADC), combining an efficient charge-sharing successive approximation register (SAR) first stage and a compact single-slope (SS) second stage. Application in large ultrasound imaging arrays is facilitated by directly interfacing the ADC with a charge-domain subarray beamformer, locally calibrating interstage gain errors and generating the SAR reference using a power-efficient local reference generator. Additional hardware-sharing between neighboring channels ultimately leads to the lowest reported area and power consumption across miniature ultrasound probe ADCs. A pitch-matched design is further enabled by an efficient split between the core circuitry and a periphery block, the latter including a datalink performing clock data recovery (CDR) and time-division multiplexing (TDM), which leads to a 12-fold total channel count reduction. A prototype of $8{\times }9$ elements was fabricated in a TSMC 0.18- $\mu \text{m}$ HV BCD technology and a 2-D PZT transducer matrix with a pitch of $160 \mu \text{m}$ , and a center frequency of 6 MHz was manufactured on the chip. The imaging device operates at up to 1000 volumes/s, generates 65-V transmit pulses, and has a receive power consumption of only 1.23 mW/element. The functionality has been demonstrated electrically as well as in acoustic and imaging experiments.

Files

A_Pitch_Matched_Transceiver_AS... (pdf)
(pdf | 4.94 Mb)
- Embargo expired in 01-07-2023
License info not available