24096-Element CMUT Array" (IEEE Journal of Solid-State Circuits (2025) 60:4 (1397-1410) DOI: 10.1109/JSSC.2025.3534087)">
IN [1], there is a mistake in the timing diagram shown in Fig. 6. Switches S 1-S 4 are skipping some of the samples and the rate at which they are operating implies a TDM rate of 10 MHz, whereas (as described in [1]) this should be 20 MHz. In the updated Fig. 6, S 1-S 4 have been updated and a minor change has been made to the timing shown for switches Q1 and Q2, such that the correct TDM rate is indicated and no sample provided to the S/H stage via N1-N4 is skipped in the diagram. (Figure presented). ...
This article presents a 4096-element ultrasound probe for high volume-rate (HVR) cardiovascular imaging. The probe consists of two application-specific integrated circuits (ASICs), each of which interfaces with a 2048-element monolithically-integrated capacitive micro-machined ultrasound transducer (CMUT) array. The probe can image a 60° × 60° × 10-cm volume at 2000 volumes/s, the highest volume-rate with in-probe channel-count reduction reported to date. It uses 2 × 2 delay-and-sum micro-beamforming (μBF) and 2× time-division multiplexing (TDM) to achieve an 8× receive (RX) channel-count reduction. Equalization, trained using a pseudorandom bit-sequence generated on the chip, reduces TDM-induced crosstalk by 10 dB, enabling power-efficient scaling of the cable drivers. The ASICs also implement a novel transmit (TX) beamformer (BF) that operates as a programmable digital pipeline, which enables steering of arbitrary pulse-density modulated (PDM) waveforms. The TX BF drives element-level 65 V unipolar pulsers, which in turn drive the CMUT array. Both the TX BF and RX μBF are programmed with shift-registers (SRs) that can either be programmed in a row-column fashion for fast upload times, or daisy-chain fashion for a higher flexibility. The layout of the ASICs is matched to the 365-μm-pitch monolithically-integrated CMUT array. While operating, the RX and logic power consumption per element is 0.85 and 0.10 mW, respectively. TX power consumption is highly waveform dependent, but is nominally 0.34 mW. Compared to the prior art, the probe has the highest volume rate, and features among the largest imaging arrays (both in terms of element-count and aperture) with a high flexibility in defining the TX waveform. These properties make it a suitable option for applications requiring HVR imaging of a large region of interest. ...
Book chapter (2024) - Michiel Pertijs, Yannick Hopf, Peng Guo
While medical imaging using ultrasound is an established field, technical advances are enabling a range of new-use cases and associated new ultrasound imaging devices. Examples include catheters capable of providing real-time 3D images to guide minimally invasive interventions and wearable devices for new monitoring and diagnostic applications. In contrast with conventional probes, which contain little or no electronics, these new devices need to become “smart”: integrated circuits need to be integrated into the probe to interface in a pitch-matched fashion with the many transducer elements (typically 1000+) needed for real-time 3D imaging. This chapter discusses the challenges associated with the design of such pitch-matched integrated circuits, focusing on strategies for channel-count reduction, beamforming, and digitization. The chapter includes a case study of a state-of-the-art catheter-based design for high-frame-rate 3D intracardiac imaging. ...
This article presents an application-specific integrated circuit (ASIC) for catheter-based 3-D ultrasound imaging probes. The pitch-matched design implements a comprehensive architecture with high-voltage (HV) transmitters, analog front ends, hybrid beamforming analog-To-digital converters (ADCs), and data transmission to the imaging system. To reduce the number of cables in the catheter while maintaining a small footprint per element, transmission (TX) beamforming is realized on the chip with a combination of a shift register (SR) and a row/column (R/C) approach. To explore an additional cable-count reduction in the receiver part of the design, a channel with a combination of time-division multiplexing (TDM), subarray beamforming, and multi-level pulse amplitude modulation (PAM) data transmission is also included. This achieves an 18-fold cable-count reduction and minimizes the power consumption in the catheter by a load modulation (LM) cable driver. It is further explored how common-mode interference can limit beamforming gain and a strategy to reduce its impact with local regulators is discussed. The chip was fabricated in TSMC 0.18-m HV BCD technology and a 2-D PZT transducer matrix of 16 × 18 elements with a pitch of 160 m and a center frequency of 6 MHz was manufactured on the chip. The system can generate all required TX patterns at up to 30 V, provides quick settling after the TX phase, and has an reception (RX) power consumption of only 1.12 mW/element. The functionality and operation of up to 1000 volumes/s have been demonstrated in electrical and acoustic imaging experiments. ...
Objective: Described here is the development of an ultrasound matrix transducer prototype for high-frame-rate 3-D intra-cardiac echocardiography. Methods: The matrix array consists of 16 × 18 lead zirconate titanate elements with a pitch of 160 µm × 160 µm built on top of an application-specific integrated circuit that generates transmission signals and digitizes the received signals. To reduce the number of cables in the catheter to a feasible number, we implement subarray beamforming and digitization in receive and use a combination of time-division multiplexing and pulse amplitude modulation data transmission, achieving an 18-fold reduction. The proposed imaging scheme employs seven fan-shaped diverging transmit beams operating at a pulse repetition frequency of 7.7 kHz to obtain a high frame rate. The performance of the prototype is characterized, and its functionality is fully verified. Results: The transducer exhibits a transmit efficiency of 28 Pa/V at 5 cm per element and a bandwidth of 60% in transmission. In receive, a dynamic range of 80 dB is measured with a minimum detectable pressure of 10 Pa per element. The element yield of the prototype is 98%, indicating the efficacy of the manufacturing process. The transducer is capable of imaging at a frame rate of up to 1000 volumes/s and is intended to cover a volume of 70° × 70° × 10 cm. Conclusion: These advanced imaging capabilities have the potential to support complex interventional procedures and enable full-volumetric flow, tissue, and electromechanical wave tracking in the heart. ...
Poster (2023) - D. Santos, Y. Hopf, B. Ossenkoppele, J. Bosch, R. Vos, M. Pertijs, N. de Jong, M. Verweij
High frame rate 3D intracardiac echography (3D-ICE) might enable full-volumetric flow, tissue, and electro-mechanical wave tracking in the heart, thus supporting complex interventional procedures. Designing ICE transducers is very challenging due to the constraints imposed by the catheter's diameter on the transducer aperture and cable accommodation within the shaft. Current ICE probes only offer a 2D or limited 3D field of view at low frame rates. This work presents the design, fabrication, and characterization of an ICE prototype transducer aiming at high frame rate 3D imaging with a large field of view. ...
Poster (2023) - N. Radeljic-Jakic, A. Flikweert, Y. Hopf, N. Rozsa, M. Pertijs
In traditional 2-D ultrasound probes, a 1-D transducer array is directly connected to an imaging system. With the introduction of 3-D probes that have 2-D arrays with thousands of elements, this approach has become impractical. Ultrasound ASICs can enable this transition by shifting part of the system functionality into the probe to reduce interconnect and cost. On-chip implementation of the analog-to-digital converter (ADC) has recently been shown to be particularly beneficial but comes with a significant power and area penalty. Current ultrasound converters are commonly implemented as successive approximation register (SAR) ADCs and designed following general-purpose design methodologies. In this work, the impact of SAR ADC non-idealities on postprocessed images is studied to achieve better trade-offs between performance and cost for ultrasound imaging. ...
Doctoral thesis (2023) - Y.M. Hopf
This thesis describes the design of application-specific integrated circuits (ASICs) for high-frame-rate 3D intracardiac echocardiography (ICE) probes. The work follows the whole process from the analysis of the application to the testing of a prototype with all envisioned functionality. It starts with the evaluation of the main challenges posed by ultrasound imaging from the tip of a catheter with a diameter in the order of 3 mm and its implications on the circuit design. A broad literature review in the field of ICE probe development but also regarding general ultrasound electronics provides the basis for this work and motivates later design choices. A high-frame-rate 3D imaging scheme is developed based on the needs of the application and leads to the design targets of the imager array and electronics. The final ASIC architecture is reached over the course of two prototypes to enable the evaluation of block-level functionality and advanced debugging. The thesis provides a detailed description of the circuit implementation with introduced novelties and shows characterization results in the electrical as well as acoustic domain. In the following, the contents of each chapter of the main matter are summarized in one paragraph.... ...
Journal article (2022) - Mehdi Soozande, Boudewine W. Ossenkoppele, Yannick Hopf, Michiel A.P. Pertijs, Martin D. Verweij, Nico De Jong, Hendrik J. Vos, Johan G. Bosch
Atrial fibrillation (AF) is the most common cardiac arrhythmia and is normally treated by RF ablation. Intracardiac echography (ICE) is widely employed during RF ablation procedures to guide the electrophysiologist in navigating the ablation catheter, although only 2-D probes are currently clinically used. A 3-D ICE catheter would not only improve visualization of the atrium and ablation catheter, but it might also provide the 3-D mapping of the electromechanical wave (EW) propagation pattern, which represents the mechanical response of cardiac tissue to electrical activity. The detection of this EW needs 3-D high-frame-rate imaging, which is generally only realizable in tradeoff with channel count and image quality. In this simulation-based study, we propose a high volume rate imaging scheme for a 3-D ICE probe design that employs 1-D micro-beamforming in the elevation direction. Such a probe can achieve a high frame rate while reducing the channel count sufficiently for realization in a 10-Fr catheter. To suppress the grating-lobe (GL) artifacts associated with micro-beamforming in the elevation direction, a limited number of fan-shaped beams with a wide azimuthal and narrow elevational opening angle are sequentially steered to insonify slices of the region of interest. An angular weighted averaging of reconstructed subvolumes further reduces the GL artifacts. We optimize the transmit beam divergence and central frequency based on the required image quality for EW imaging (EWI). Numerical simulation results show that a set of seven fan-shaped transmission beams can provide a frame rate of 1000 Hz and a sufficient spatial resolution to visualize the EW propagation on a large 3-D surface. ...
Journal article (2022) - Yannick M. Hopf, Boudewine W. Ossenkoppele, Michiel A.P. Pertijs, Mehdi Soozande, Emile Noothout, Zu Yao Chang, Chao Chen, Hendrik J. Vos, Johan G. Bosch, Martin D. Verweij, Nico de Jong
In this article, an application-specific integrated circuit (ASIC) for 3-D, high-frame-rate ultrasound imaging probes is presented. The design is the first to combine element-level, high-voltage (HV) transmitters and analog front-ends, subarray beamforming, and in-probe digitization in a scalable fashion for catheter-based probes. The integration challenge is met by a hybrid analog-to-digital converter (ADC), combining an efficient charge-sharing successive approximation register (SAR) first stage and a compact single-slope (SS) second stage. Application in large ultrasound imaging arrays is facilitated by directly interfacing the ADC with a charge-domain subarray beamformer, locally calibrating interstage gain errors and generating the SAR reference using a power-efficient local reference generator. Additional hardware-sharing between neighboring channels ultimately leads to the lowest reported area and power consumption across miniature ultrasound probe ADCs. A pitch-matched design is further enabled by an efficient split between the core circuitry and a periphery block, the latter including a datalink performing clock data recovery (CDR) and time-division multiplexing (TDM), which leads to a 12-fold total channel count reduction. A prototype of $8{\times }9$ elements was fabricated in a TSMC 0.18- $\mu \text{m}$ HV BCD technology and a 2-D PZT transducer matrix with a pitch of $160 \mu \text{m}$ , and a center frequency of 6 MHz was manufactured on the chip. The imaging device operates at up to 1000 volumes/s, generates 65-V transmit pulses, and has a receive power consumption of only 1.23 mW/element. The functionality has been demonstrated electrically as well as in acoustic and imaging experiments. ...

In this letter, a compact high-voltage (HV) transmit circuit for dense 2-D transducer arrays used in 3-D ultrasonic imaging systems is presented. Stringent area requirements are addressed by a unipolar pulser with embedded transmit/receive switch. Combined with a capacitive HV level shifter, it forms the ultrasonic HV transmit circuit with the lowest reported HV transistor count and area without any static power consumption. The balanced latched-based level shifter implementation makes the design insensitive to transients on the HV supply caused by pulsing, facilitating application in probes with limited local supply decoupling, such as imaging catheters. Favorable scaling through resource sharing benefits massively arrayed architectures while preserving full individual functionality. A prototype of 8 × 9 elements was fabricated in the TSMC 0.18 μm HV BCD technology and a 160μm×160μm PZT transducer matrix is manufactured on the chip. The system is designed to drive 65-V peak-to-peak pulses on 2-pF transducer capacitance and hardware sharing of six elements allows for an area of only 0.008 mm2 per element. Electrical characterization as well as acoustic results obtained with the 6-MHz central frequency transducer are demonstrated. ...

Intra-cardiac echography (ICE) probes (Fig. 32.2.1) are widely used in electrophysiology for their good procedure guidance and relatively safe application. ASICs are increasingly employed in these miniature probes to enhance signal quality and reduce the number of connections needed in mm-diameter catheters [1]-[5]. 3D visualization in real-time is additionally enabled by 2D transducer arrays with, for each transducer element, a high-voltage (HV) transmit (TX) part, to generate acoustic pulses of sufficient pressure, and a receive (RX) path, to process the resulting echoes. To achieve the required reduction in RX channels, micro-beamforming (BF), which merges the signals from a subarray using a delay-and-sum operation, has been shown to be an effective solution [3], [4]. However, due to the frame-rate reduction that is associated with BF, these designs cannot serve emerging high-frame-rate imaging modes (1000 volumes/s) like 3D blood-flow and elastography imaging. In-probe digitization has recently been investigated to provide further channel-count reduction, make data transmission more robust, and enable pre-processing in the probe [1]-[3]. However, these earlier designs have either no TX functionality [2], [3] or only low-voltage (LV) TX [1] integrated. Combining BF and digitization with area-hungry HV transmitters in a pitch-matched scalable fashion while supporting high-frame-rate imaging remains an unmet challenge. The work presented in this paper meets this target, enabled by a hybrid ADC, the small die size of which allows for co-integration with 65V element-level pulsers. ...
This work describes an ASIC design for high-frame-rate 3D intracardiac echocardiography probes. The chip is the first to combine element-level high-voltage pulsers and time-gain-compensation analog frontends as well as subarray beamformers and in-probe digitization in a pitch-matched fashion. The integration challenge is met by a shared hybrid beamforming ADC with the highest reported area and power efficiency. The achieved beamformer size of three elements enables acquisition at 1000 volumes/s while, in combination with a custom datalink, still providing sufficient channel-count reduction for catheter integration. ...