Correction to "A 2000-Volumes/s 3-D Ultrasound Probe With Monolithically-Integrated 23 × 23-mm24096-Element CMUT Array" (IEEE Journal of Solid-State Circuits (2025) 60:4 (1397-1410) DOI: 10.1109/JSSC.2025.3534087)
Nuriel N.M. Rozsa (TU Delft - Electronic Instrumentation)
Taehoon Kim (Student TU Delft)
Peng Guo (TU Delft - Bio-Electronics)
Yannick M. Hopf (TU Delft - Electronic Instrumentation, Sonosilicon)
Jason Voorneveld (Erasmus MC)
Djalma Simoes Dos Santos (TU Delft - ImPhys/Medical Imaging, TU Delft - ImPhys/Verweij group)
Emile Noothout (TU Delft - ImPhys/Verweij group)
Zu Yao Chang (TU Delft - Electronic Instrumentation)
Chao Chen (TU Delft - Transport and Logistics, SonoSilicon)
Nico De Jong (TU Delft - ImPhys/Verweij group, Erasmus MC)
Hendrik J. Vos (TU Delft - ImPhys/Verweij group, Erasmus MC)
Johan G. Bosch (Erasmus MC)
Martin D. Verweij (TU Delft - ImPhys/Medical Imaging, TU Delft - ImPhys/Verweij group)
Michiel A.P. Pertijs (TU Delft - Electronic Instrumentation)
undefined More Authors (External organisation)
More Info
expand_more
Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons.
Abstract
IN [1], there is a mistake in the timing diagram shown in Fig. 6. Switches S 1-S 4 are skipping some of the samples and the rate at which they are operating implies a TDM rate of 10 MHz, whereas (as described in [1]) this should be 20 MHz. In the updated Fig. 6, S 1-S 4 have been updated and a minor change has been made to the timing shown for switches Q1 and Q2, such that the correct TDM rate is indicated and no sample provided to the S/H stage via N1-N4 is skipped in the diagram. (Figure presented).