97 records found
1
The long-term revision risk of RSA-tested and non-RSA-tested TKA implants in The Netherlands
A Dutch arthroplasty register study
Background and purpose — Radiostereometric analysis (RSA) of total knee arthroplasty (TKA) is used as an early safeguard during the phased evidence-based introduction of new implants. The goal of our study was to compare the long-term revision risk between RSA-tested implants and non-RSA-tested implants in the Netherlands using patient-level data. Methods — All primary TKAs between 2007 and 2016 from the Dutch Arthroplasty Register were included, and procedures with an RSA-tested implant were identified. Both all-cause major revision risk and revision risk because of loosening were calculated at 5 and 10 years postoperatively using Kaplan–Meier analyses. Sensitivity analyses were per-formed with stricter definitions of implant characteristics to classify procedures as RSA-tested, to avoid camouflage of different subdesigns within the same brand implant portfolio. Results — 83,638 RSA-tested and 104,105 non-RSA-tested TKAs were included. Cumulative all-cause major revision percentages for the RSA-tested group at 5 and 10 years were 2.2% (95% confidence interval [CI] 2.1–2.3) and 3.6% (CI 3.4–3.7), respectively, compared with 2.5% (CI 2.4–2.6) and 3.3% (CI 3.2–3.4) for the non-RSA-tested group. RSA-tested TKAs showed higher 10-year revision risks because of loosening than non-RSA-tested procedures (1.8%, CI 1.7– 1.9 vs 1.4%, CI 1.3–1.4, respectively). Comparable results were found after stratification by various patient characteristics and with stricter classification approaches. Conclusion — Regarding all-cause revision risk, RSA-tested TKAs had a slightly lower risk at 5 years. However, at 10 years the RSA-tested TKAs had a higher all-cause revision risk and higher revision risk because of loosening compared with the non-RSA-tested TKAs.
Minimally-invasive re-stabilisation of loosened hip implants using percutaneous periprosthetic cement injections
A feasible option to reduce pain in frail patients
Introduction: Loosening of hip implants causes severe pain and disability. Revision surgery, the standard treatment, carries high risks, especially for frail patients. Therefore, a less-invasive treatment option is needed for patients with incapacitating pain. We evaluated the safety and efficacy of percutaneous periprosthetic cement injection (PPCI) to restabilise loosened hip implants. Methods: All patients treated with PPCIs of hip implants between 2004 and 2020 were retrospectively included (median follow-up 24 months, IQR 11–51). Cement was injected in the periprosthetic space of the loosened hip implant under fluoroscopy. Patient characteristics and the Charlson Comorbidity Index (CCI) were measured. Pain reduction was measured by the Harris Hip pain-subscore and adverse events were recorded. Results: 34 patients were included (mean age 83 years, median CCI 6 points). 10 patients had a second PPCI after 12 months (median, IQR 7–27). 2 patients needed revision surgery. 30 patients had pain reduction and 4 had no pain reduction. Adverse events included femoral fissure during needle placement (n = 1) and progressive migration of a loosened acetabular cup (despite the cement injection; n = 2). Conclusions: PPCIs for loosened hip implants is feasible in frail patients with pain reduction in the majority of patients and has limited adverse events.
Background and purpose — Pooling data on the performance of total knee (TK) implants across registries is only possible if the same TK implant is used across multiple registries and if used in patients with similar characteristics. We assessed to what extent specific TK implants: (i) are used across multiple registries or only in a single registry; and (ii) differ in patient characteristics between registries. Methods — All primary TK implants implanted between January 2020 and December 2021 in the Danish, Dutch, German, and Italian registries were included. We determined the number of registries using a specific TK implant (based on combined femoral-tibial component brand name and fixation/congruency/mobile bearing insert/patella usage). Patient characteristics (age/body mass index [BMI]/sex/ diagnosis osteoarthritis) were compared across registries for TK implants used in ≥ 2 registries ≥ 100 times. Results — 813 different TK implants (577,351 procedures) were used across the 4 registries, of which 53 TK implants (7%) were used in 1 registry (8,000 procedures). 760 different TK implants (569,351 procedures; 99%) were used in ≥ 2 registries of which 47 different TK implants (393,954 procedures; 68%) were used in ≥ 2 registries and ≥ 100 times. Statistically and clinically significant differences in age for the same TK implant across registries were observed for 29 TK implants (62%) and 3 TK implants (6%), respectively; for other characteristics these were for BMI 30 (64%) and 0 (0%) TK implants; for male proportion 23 (49%) and 17 (36%) TK implants; and for diagnosis of osteoarthritis 42 (89%) and 34 (72%) TK implants, respectively. Conclusion — Most specific TK implants and TK procedures were used across multiple registries, but they were often used in patients with different characteristics. This has an impact on comparing implant performances between registries.
Before a high-risk medical device is approved for implantation into patients, there should be evidence not only of its performance and safety with a favourable benefit-risk ratio, but also of its clinical efficacy. Regulatory guidance on study methodologies is lacking, however, so the European Commission funded the CORE–MD project (Coordinating Research and Evidence for Medical Devices) to advise regulators on appropriate designs for clinical trials of high-risk devices. The CORE–MD consortium recommends that evaluation should be planned in four stages. Randomised controlled trials should be performed more often, against active comparators reflecting the best available treatment, or using sham interventions with ethical safeguards. Large trials can be managed efficiently using an electronic database or registry. Non-randomised clinical studies can apply objective performance criteria or other validated patient-relevant outcome measures, with adjustments to minimise bias. Full transparency of results from clinical investigations is essential. Proportionate regulation of breakthrough or orphan devices for independently-defined serious unmet needs may involve approval with less evidence, but on condition of subsequent confirmatory studies. These CORE–MD consensus proposals have been submitted to European Union medical device regulators, to be considered as a basis for more transparent and predictable requirements for clinical evidence. Funding: The CORE–MD project was funded as a Coordination and Support action from the European Union Horizon 2020 research and innovation programme, under grant agreement 965246.
Background and purpose — The safety and performance of hip and knee prostheses can be assessed by ana-lyzing peer-reviewed literature, registry reports, and safety notices published by national competent authorities/regula-tory agencies, or manufacturers. The percentage of hip and knee prostheses with a safety signal published through any of these data sources is unknown. We aimed to assess the frequency of signals identified for a random sample of 10 hip stems, 10 hip cups, and 10 knee implants. Methods — 3 literature libraries were searched to find safety signals defined as information on patterns/occurrences that may alter the device’s benefit–risk profile, reported in peer-reviewed publications for the randomly selected implants. Annual registry reports from 5 national registries were examined to check whether any of the selected implants had outlier performance. The CORE-MD post-market surveillance (PMS) tool was used to collect all related safety notices from 13 competent authority/regulatory agency websites. Manufacturers’ websites were screened for any reported safety information. Results — Safety signals were identified for 21 of the 30 randomly selected implants: 18 identified by registries, 7 by the CORE-MD PMS tool, and 8 based on literature, with 10 implants identified by multiple sources. There was no systematic pattern in timing of publication with a particular source publishing safety signals earlier than other sources. Conclusion — 70% of the randomly selected hip and knee prostheses had ≥ 1 safety signals published, with registries as the source for the majority. No single source identified all 21 implants with signals, which highlights the need for a comprehensive surveillance strategy to aggregate safety signals from multiple sources.
Objectives Medical device registries in Europe report limited information about their structure and methodological characteristics. This hinders their utility for evaluation of medical device safety and performance under the Medical Device Regulation. This study aimed to define a minimum checklist of items necessary for regulators to assess the quality of evidence produced using registry data for the evaluation of medical device safety and performance. Design A three-round Delphi panel. Setting A task within the Coordinating Research and Evidence for Medical Devices project. Participants 101 experts in the medical device community (healthcare professionals, methodologists, registry experts, regulators, and assessors from notified bodies) were invited. Interventions Based on a literature review and expert advice, 27 items relating to the quality of registry data and the analysis of medical device safety and performance were selected. In round 1, participants selected which items were required for a minimum checklist. They could also propose new items. Items selected by ≥70% of participants indicated consensus. Remaining items were discussed in round 2, resulting in a final checklist that was ranked by participants for importance (round 3). Main outcome measures Consensus of items to be included in the minimum checklist. Results 51 experts participated in round 1, achieving consensus on 18 (67%) items and suggesting 12 items. After discussion in round 2, 5 additional items were selected, resulting in a final set of 15 data quality items and 8 data analysis items. The most important items were ‘completeness of procedures’ (data quality) and ‘definition of outcome analyzed’” (quality of analysis). Conclusions Reporting all items from the minimum checklist will facilitate judgment of the utility of registry data to evaluate medical devices during post-market surveillance.
Cumulative risk of revision after primary total hip arthroplasty in registries internationally
Systematic review and meta-analysis of selected hip stems and cups
Purpose: The objective was to investigate the consistency in cumulative revision rates (CRRs) for a selection of total hip arthroplasty cups and stems across national/regional hip arthroplasty registries worldwide. • Methods: Ten cups and ten stems from total hip systems were randomly selected. Two frequently used implants across registries were added, totalling 11 cups and 11 stems. CRRs and 95% CIs were extracted from the latest annual registry reports using these implants. CRRs were pooled for each cup or stem, and differences between cup-stem combinations and between registries were investigated. • Results: CRRs were available for ten cups and eight stems from eight registries, totalling 552,148 cups and 727,447 stems. Follow-up was 1–20 years. The 5-year CRR pooled for all cups was 2.9% (95% CI: 2.3–3.6) and for all stems, 3.0% (95% CI: 2.4–3.8). Homogeneous (consistent) CRRs with respect to both associated implant and country were observed for two cups and three stems. Significant differences in CRR were identified in one cup by associated implant only, in one cup by registry only, and in two cups and four stems for both. Sparse data prevented evaluation of four cups and one stem. • Conclusion: Registries’ annual reports provide a large amount of publicly available information on CRRs of specific implants. These CRRs can be synthesised to improve the assessment of implant performance over time. Our CRR analysis represents a promising approach to detect implants with a consistent low- or high-risk pattern across registries.
Acetabular defects pose significant challenges in orthopedic surgery, particularly in revision total hip arthroplasty (THA). Here, we design, additively manufacture, and evaluate shape-morphing porous implants with kinematic structures to address these defects. Three defect types were examined using synthetic hemipelvis models: posterior wall, cranial-posterior combination, and central-posterior defects. The implants were secured with screws and bone cement, and their surface conformity was assessed through micro computed tomography (µCT). Biomechanical performance was evaluated under quasi-static compression and cyclic loading conditions. Results demonstrated high surface conformity of the flexible mesh across all defect types, with minimal differences from healthy acetabula (< 10 mm). The mesh implants exhibited strong load-bearing capacity, with failures occurring only in the pubic region of the hemipelvis, while both the implants and mesh-cement interfaces remained intact. The implants withstood cyclic loading simulating half the body weight of a 80 kg patient for >1000,000 loading cycles with no evidence of fatigue failure, further confirming their durability. These findings suggest that the flexible mesh implant provides a potential solution for complex acetabular defects, offering anatomical conformity and mechanical stability, even in cases where conventional mesh grafts might be inadequate. Future studies, including cadaveric testing and clinical trials, are necessary to further validate these results in (pre-)clinical settings. Statement of significance: This study addresses the need for adaptable solutions to complex acetabular defects in revision total hip arthroplasty (THA). Traditional implants struggle to conform to severe bone loss and irregular geometries, risking suboptimal fit, and implant migration. We introduce a 3D-printed, shape-morphing porous implant with kinematic structures, offering high anatomical conformity, mechanical robustness, and support for bone graft integration. Combining the adaptability of patient-specific implants with the efficiency of standard designs, this implant reduces lead times while enabling a tailored fit. This innovative approach provides a reliable solution for managing complex defects, addressing limitations of conventional implants, and improving outcomes in orthopedic reconstruction.
Arthroplasty registries at a glance
An initiative of the International Society of Arthroplasty Registries (ISAR) to facilitate access, understanding, and reporting of registry data from an international perspective
Background and purpose — The amount of information publicly available from arthroplasty registries is large but could be used more effectively. This project aims to improve the knowledge concerning existing registries to facilitate access, transparency, harmonization, and reporting. Methods — Within the International Society of Arthroplasty Registries (ISAR) we aimed at developing, testing, adopting, and making publicly available a short, standardized registry description with items considered relevant for stakeholders using a cross-sectional study survey. Items were chosen based on a literature review and expert advice, selected by 9 ISAR working group members, tested iteratively in 3 registries, and commented upon by 4 external experts. All 29 ISAR member registries as of July 2023 were invited to participate in the project. Results — Included items covered general descriptive information regarding registries, information related to governance, outcomes, data quality, data access, and registry production. The template was adopted, completed, and made publicly available by 25 of the 29 registries. Of those, 2/3 were national registries. 23 captured both hip and knee arthroplasties and 10 captured shoulder arthroplasties. Most registries had public reporting of data quality, methods, and results. Data was accessible in all but 2 registries, mainly as aggregated data. Important items relevant to registry quality for researchers to consistently indicate in scientific papers include scope, inclusion criteria, outcomes definitions, coverage/completeness, and validation processes. Conclusion — This ISAR initiative implemented a short, standardized description to facilitate appropriate use of orthopedic registry data worldwide relevant for a diverse group of stakeholders including researchers, industry, public health and regulatory agencies.
Background: Orthopaedic Data Evaluation Panel (ODEP) ratings of total hip (TH) and total knee (TK) implants are informative for assessing implant performance. However, the validity of ODEP ratings across multiple registries is unknown. Therefore, we aimed to assess, across multiple registries, whether TH and TK implants with a higher ODEP rating (i.e., an A* rating) have lower cumulative revision risks (CRRs) than those with a lower ODEP rating (i.e., an A rating) and the extent to which A* and A-rated implants would be A*-rated on the basis of the pooled registries’ CRR. Methods: Implant-specific CRRs at 3, 5, and 10 years that were reported by registries were matched to ODEP ratings on the basis of the implant name. A meta-analysis with random-effects models was utilized for pooling the CRRs. ODEP benchmark criteria were utilized to classify these pooled CRRs. Results: A total of 313 TH cups (54%), 356 TH stems (58%), 218 TH cup-stem combinations (34%), and 68 TK implants (13%) with unique brand names reported by registries were matched to an ODEP rating. Given the low percentage that matched, TK implants were not further analyzed. ODEP-matched TH implants had lower CRRs than TH implants without an ODEP rating at all follow-up time points, although the difference for TH stems was not significant at 5 years. No overall differences in CRRs were found between A* and A-rated TH implants, with the exception of TH cup-stem combinations, which demonstrated a significantly lower CRR for A*A*-rated cup-stem combinations at the 3-year time point. Thirty-nine percent of A*-rated cups and 42% of A*-rated stems would receive an A* rating on the basis of the pooled registries’ CRR at 3 years; however, 24% of A-rated cups and 31% of A-rated stems would also receive an A* rating, with similar findings demonstrated at longer follow-up. Conclusions: At all follow-up time points, ODEP-matched TH implants had lower CRRs than TH implants without an ODEP rating. Given that the performance of TH implants varied across countries, registries should first validate ODEP ratings with use of country-specific revision data to better guide implant selection in their country. Data source transparency and the use of revision data from multiple registries would strengthen the ODEP benchmarks.
Safety notices and registry outlier data measure different aspects of safety and performance of total knee implants
A comparative study of safety notices and register outliers
Background and purpose — Safety notices for medical devices such as total knee arthroplasty (TKA) implants may indicate problems in their design or performance that require corrective action to prevent patient harm. Safety notices are often published on national Ministries of Health or regulatory agencies websites. It is unknown whether problems triggering safety notices identify the same implants as those identified by registries as “outlier.” We aimed to assess the extent to which safety notices and outlier identification in registries signal the same or different TKA implants. Methods — The CORE-MD tool, an automated web scraper tool, was used to collect safety notices related to TKA implants on 13 national Ministries of Health websites and regulatory agencies. Safety notices were defined accord-ing to the Medical Device Regulation (MDR) as “a com-munication sent by a manufacturer to users or customers in relation to a field safety corrective action.” Identified TKA outliers, defined as having a significantly higher revision risk than other comparable TKA implants, were extracted from registry reports. Results — 787 safety notices for 38 TKA implants and 35 TKA outliers were identified, together identifying 47 unique TKA implants. 26 (55%) TKA implants had safety notices and were also outliers, 12 (26%) TKA implants had only safety notices, and 9 (19%) were outliers only. TKA implants with safety notices only had similar types of problems to TKA outliers with safety notices, with “Manufac-turing/Packaging/Shipping” problems being most frequent (44%). Cumulative revision risks (1/5/10 years) were lower for TKA implants with safety notices only than for TKA out-liers with safety notices. Conclusion — 55% of the TKA with a safety notice were identified as outliers in the registry, whereas around 25% of TKA outliers are not the subject of publicly released safety notices, with safety notices pointing to TKA implants not identified by registries as potentially having a higher risk of failure. This suggests that safety notices and registry outlier data measure different aspects of safety and performance.
Methods: PubMed, Embase, Web of Science, Cochrane Library, and Academic Search Premier were systematically searched up to January 2020 for studies including THA/TKA and biomarkers to assess loosening. Two reviewers independently screened records, extracted data, and assessed the risk of bias using the ICROMS tool to classify the quality of the studies.Results: Twenty-eight (three high-quality) studies were included, reporting on a median of 48 patients (interquartile range 28–69). Serum and urine markers were evaluated in 22 and 10 studies, respectively. Tumor necrosis factor α and osteocalcin were significantly higher in loosened compared with stable implants. Urinary N-terminal telopeptide had significantly elevated levels in loosened prostheses.Conclusion: Several serum and urine markers were promising in discriminating between loosened and stable implants. We recommend future studies to evaluate these biomarkers in a longitudinal fashion to assess whether progression of loosening is associated with a change in these biomarkers. In particular, high-quality studies assessing the usability of these biomarkers are needed. ...
Background: Loosening is a major cause for failure of total hip and total knee arthroplasties (THAs/TKAs). Preemptive diagnostics of asymptomatic loosening could open strategies to prevent gross loosening. A multitude of biomarkers may discriminate between loosened and stable implants, but it is unknown which have the best performance. The present systematic review aimed to assess which biomarkers have shown the most promising results in discriminating between stable and aseptic loosened THAs and TKAs.
Methods: PubMed, Embase, Web of Science, Cochrane Library, and Academic Search Premier were systematically searched up to January 2020 for studies including THA/TKA and biomarkers to assess loosening. Two reviewers independently screened records, extracted data, and assessed the risk of bias using the ICROMS tool to classify the quality of the studies.
Results: Twenty-eight (three high-quality) studies were included, reporting on a median of 48 patients (interquartile range 28–69). Serum and urine markers were evaluated in 22 and 10 studies, respectively. Tumor necrosis factor α and osteocalcin were significantly higher in loosened compared with stable implants. Urinary N-terminal telopeptide had significantly elevated levels in loosened prostheses.
Conclusion: Several serum and urine markers were promising in discriminating between loosened and stable implants. We recommend future studies to evaluate these biomarkers in a longitudinal fashion to assess whether progression of loosening is associated with a change in these biomarkers. In particular, high-quality studies assessing the usability of these biomarkers are needed.
The use of MRI and arthroscopy are considered low-value care in most patients with degenerative knee disease. To reduce these modalities, there have been multiple efforts to increase awareness. Reductions have been shown for general hospitals (GH), but it is unclear whether this may be partly explained by a shift of patients receiving these modalities in independent treatment centers (ITCs). The aims of this study were to assess (i) whether the trend in use of MRI and arthroscopy in patients with degenerative knee disease differs between ITCs and GH, and (ii) whether the Dutch efforts to raise awareness on these recommendations were associated with a change in the trend for both types of providers. All patients insured by a Dutch healthcare insurer aged ≥50 years with a degenerative knee disease who were treated in a GH or ITC between July 2014 and December 2019 were included. Linear regression was used with the quarterly percentage of patients receiving an MRI or knee arthroscopy weighted by center volume, as the primary outcome. Interrupted time-series analysis was used to evaluate the effect of the Dutch efforts to raise awareness. A total of 14 702 patients included were treated in 90 GHs (n = 13 303, 90.5%) and 29 ITCs (n = 1399, 9.5%). Across the study period, ITCs on an average had a 16% higher MRI use (P <. 001) and 9% higher arthroscopy use (P =. 003). MRI use did not change in both provider types, but arthroscopy use significantly decreased and became stronger in ITCs (P =. 01). The Dutch efforts to increase awareness did not significantly influence either MRI or arthroscopy use in ITCs (P =. 55 and P =. 84) and GHs (P =. 13 and P =. 70). MRI and arthroscopy uses were higher in ITCs than GHs. MRI use did not change significantly among patients ≥ 50 years with degenerative knee disease in both provider types between 2014 and 2019. MRI- and arthroscopy use decreased with ITCs on average having higher rates for both modalities, but also showing a stronger decrease in arthroscopy use. The Dutch efforts to increase awareness did not accelerate the already declining trend in the Netherlands.
Multiple Joint Arthroplasty in Hip and Knee Osteoarthritis Patients
A National Longitudinal Cohort Study
Background: Many patients suffer from osteoarthritis (OA) in multiple joints, possibly resulting in multiple joint arthroplasties (MJAs). Primarily, we determined the cumulative incidence (Cin) of MJA in hip and knee joints up to 10 years. Secondly, we calculated the mean time between the first and subsequent joint arthroplasty, and evaluated the different MJA trajectories. Lastly, we compared patient characteristics and outcomes (functionality and pain) after surgery between MJA patients and single hip arthroplasty or knee arthroplasty (HA and KA) patients. Methods: Primary index (first) HA or KA for OA were extracted from the Dutch Arthroplasty Register. The 1, 2, 5, and 10-year Cin (including competing risk death) of MJA, mean time intervals, and MJA-trajectories were calculated and stratified for primary index HA or KA. Sex, preoperative age, and body mass index were compared using ordinal logistic regression. Outcomes, measured preoperatively, 3, 6, and 12 months postoperatively (function: Hip Disability or Knee Injury and OA Outcome Score; Pain: Numerical Rating Scale), were compared using linear regression. Results: A total of 140,406 HA-patients and 140,268 KA-patients were included. One, 2, 5, and 10-year Cin for a second arthroplasty were respectively 8.9% [95% confidence interval (CI): 8.7 to 9.0], 14.3% [95% CI: 14.1 to 14.5], 24.0% [95% CI: 23.7 to 24.2], and 32.7% [95% CI: 32.2 to 33.1] after index HA, and 9.5% [95% CI: 9.4 to 9.7], 16.0% [95% CI: 15.9 to 16.2], 26.4% [95% CI: 26.1 to 26.6], and 35.8% [95% CI: 35.4 to 36.3] after index KA. The 10-year Cin for > 2 arthroplasties were small in both the index HA and KA groups. Time-intervals from first to second, third, and fourth arthroplasty were 26 [95% CI: 26.1 to 26.7], 47 [95% CI: 46.4 to 48.4], and 58 [95% CI: 55.4 to 61.1] months after index HA, and 26 [95% CI: 25.9 to 26.3], 52 [95% CI: 50.8 to 52.7], and 61 [95% CI: 58.3 to 63.4] months after index KA. There were 83% of the second arthroplasties placed in the contralateral cognate joint (ie, knee or hip). Differences in postoperative functionality and pain between MJAs and single HAs and KAs were small. Conclusions: The 10-year Cin showed that about one-third of patients received a second arthroplasty after approximately 2 years, with the majority in the contralateral cognate joint. Few patients received > 2 arthroplasties within 10 years. Being a women, having a higher body mass index, and being younger increased the odds of MJA. Postoperative outcomes were slightly negatively affected by MJA.
Long-term outcomes of pectoralis major transfer for scapular winging due to long thoracic nerve palsy
Results after a median follow-up of 17 years
Background: A pectoralis major (PM) transfer is a viable treatment option for patients with scapular winging due to long thoracic nerve (LTN) palsy not responding to nonsurgical management. However, the long-term outcomes remain unknown. Therefore, the purpose of this study was to evaluate the long-term outcome of shoulder function (ie, minimum follow-up of 10 years) and quality-of-life (QoL) of patients treated for scapular winging due to LTN palsy with a PM transfer. Methods: This observational cohort study included 15 patients (16 shoulders) who underwent PM transfer, using a tendoachilles allograft, between 1995 and 2012. Shoulder forward flexion and abduction were analyzed preoperatively, 1 year after surgery and at the final follow-up. SF-36 component scores (physical component summary (PCS) and mental component summary (MCS)) were used to evaluate the QoL. Results: Shoulder forward flexion and abduction measured in degrees improved from 86 (SD 14.5) and 82 (SD 33.8) preoperatively to 140 (SD 27.3) and 138 (31.3) at 1 year postoperatively. After a median follow-up of 17 years, mean shoulder functions were slightly lower than at 1 year postoperatively, but still better than preoperative function, ie, forward flexion 121 (SD 41.9) and abduction 122 (SD 44.5). The mean PCS score at the final follow-up was 41.9 (SD 9.7), and the mean MCS score was 49.9 (SD 12.5). Better shoulder function at the final follow-up was significantly associated with higher QoL in terms of PCS scores (P = .023), but not MCS scores (P = .287). Conclusion: The results of the present study indicate that PM transfer augmented with an achilles tendon allograft for scapular winging due to LTN palsy leads to functional improvements that persist in long term. These functional improvements likely translate to better QoL based on their association.
Tibial Baseplate Migration Is Not Associated with Change in Patient-Reported Outcome Measures and Clinical Scores after TKA
A Secondary Analysis of 5 Radiostereometric Analysis Studies with 10-Year Follow-up
Background:Radiostereometric analysis (RSA) provides highly accurate data about the migration of a total knee arthroplasty (TKA) component. However, patient-reported outcome measures (PROMs) reflect the patients' perspective of their functional status, pain, and overall health after TKA. The aim of this study was to evaluate the association between tibial implant migration and change in postoperative PROMs and clinical scores, using data pooled from long-Term follow-up RSA studies.Methods:Individual implant migration data were collected from 5 randomized RSA studies, including a total of 300 patients with 6 distinct TKA implant designs (all Stryker). Tibial implant migration (maximum total point motion [MTPM]) was evaluated with RSA at 3 months, 1 year, and 2, 5, 7, and 10 years postoperatively. The Knee Society Score (KSS)-Knee and KSS-Function and Knee Injury and Osteoarthritis Outcome Score (KOOS) subscales were collected in all studies at the same follow-up times. Linear mixed-effects models, with adjustment for TKA implant design and patient characteristics, were used to analyze the data. The 3-month follow-up visit was used as the baseline to assess the association between implant migration and PROMs across the 10-year follow-up.Results:No association between tibial implant migration and change in KSS-Knee (p = 0.384), KSS-Function (p = 0.737), KOOS-Symptoms (p = 0.398), KOOS-Pain (p = 0.699), KOOS-Activities of Daily Living (p = 0.205), KOOS-Sport and Recreation (p = 0.702), or KOOS-Quality of Life (p = 0.368) was found across the entire follow-up. Similar results were found when using the 2-year follow-up as the baseline, after which both cemented and uncemented implants are expected to have stabilized.Conclusions:Tibial baseplate migration was not associated with postoperative worsening in PROMs or clinical scores in patients who underwent TKA. These findings suggest that implant migration, as measured with RSA, measures a different parameter (i.e., implant-bone fixation) than PROMs (i.e., patient perception) and clinical scores. Therefore, to assess the performance and safety of TKA implant designs, RSA and PROMs cannot be used interchangeably during the postoperative follow-up of patients and evaluation of the fixation of knee implants.Level of Evidence:Prognostic Level III. See Instructions for Authors for a complete description of levels of evidence.
Influence of marker-selection method in radiostereometric analysis of total knee arthroplasty on tibial baseplate migration patterns
A secondary analysis of a randomized controlled trial with 5-year follow-up
Background and purpose — Different marker-selection methods are applied to represent implant and tibial seg-ments in radiostereometric analysis (RSA) studies of total knee arthroplasty (TKA). Either a consistent set of markers throughout subsequent RSA examinations (“consistent-marker method”) is used or all available markers at each follow-up (“all-marker method”). The aim of this secondary analysis was to compare marker-selection methods on individual and group level TKA migration results. Methods — Data from a randomized RSA study with 72 patients was included. Tibial baseplate migration was evalu-ated at 3 months, 1, 2, and 5 years postoperatively with both marker-selection methods. Additionally, migration was calculated using 5 fictive points, either plotted based on the consistent set of markers or all available markers. Results — Migration could be calculated with both marker-selection methods for 248 examinations. The same prosthesis and bone markers (n = 136), different prosthesis markers (n = 71), different bone markers (n = 21), or different prosthesis and bone markers (n = 20) were used. The mean difference in maximum total point motion (MTPM) between all examinations was 0.02 mm, 95% confidence interval –0.26 to 0.31 mm. 5 implants were classified as continuously migrating with the consistent-marker method versus 6 implants (same 5 plus one additional implant) with the all-marker method. Using fictive points, fewer implants were classified as continuously migrating in both marker-selection methods. Differences between TKA groups in mean MTPM were comparable with both marker-selection methods, also when fictive points were used. Conclusion — Estimated group differences in mean MTPM were similar between marker-selection methods, but individual migration results differed. The latter has implications when classifying implants for estimated risk of future loosening.
Background: The purpose of the study was to gain insight into how clinically relevant improvement in patient-reported outcome measure scores after total hip arthroplasty (THA) and total knee arthroplasty (TKA) may be underestimated or overestimated, we compared patient-reported outcome measure respondents and nonrespondents on their adverse event rates and assessed whether adverse event occurrence was associated with clinically relevant patient-reported outcome measure improvement from those without adverse events. Methods: All primary THAs and TKAs performed in 19 Dutch hospitals between January 2017 and December 2019 were included. The hip disability and osteoarthritis outcome score-physical function short form (HOOS-PS) and knee injury and osteoarthritis outcome score-physical function short form (KOOS-PS) were used to assess the physical function after THA and TKA, respectively. Adverse events included 1-year revision, 30-day readmission, 30-day complications, and long (ie, >75th percentile) length of stay. A clinically relevant improvement was defined as at least a 10-point decrease in HOOS-PS and 9 points in KOOS-PS scores. Associations between adverse events and clinically relevant HOOS-PS and KOOS-PS improvement were assessed using binary logistic regression models adjusted for patient characteristics and clustering of patients within hospitals. Results: There were 20,338 THA and 18,082 TKA procedures included. Adverse events occurred more frequently in HOOS-PS and KOOS-PS nonrespondents than in respondents. The THA patients experiencing revision, complications, or long length of stay were less likely to experience clinically relevant HOOS-PS improvements (odds ratios of 0.11 [0.06 to 0.20], 0.44 [0.30 to 0.63], and 0.66 [0.50 to 0.88], respectively). The TKA patients experiencing revision or long length of stay were less likely to experience clinically relevant KOOS-PS improvements (odds ratios of 0.26 [0.12 to 0.55] and 0.63 [0.50 to 0.80], respectively). Conclusions: Clinically relevant HOOS-PS and KOOS-PS improvements are likely overestimated, as nonrespondents had higher adverse event rates which were associated with lower likelihood to achieve clinically relevant HOOS-PS and KOOS-PS improvements.
Treatment of neurogenic scapular winging
A systematic review on outcomes after nonsurgical management and tendon transfer surgery