A low-cost, highly versatile, pulsed RF - pulsed I-V isothermal device characterization setup is presented. The realized setup combines a synthetic instrument high dynamic range pulsed network analyzer with pulsed I-V measurements. The resulting configuration facilitates very accurate characterization of low-power as well as high-power devices over a wide range of bias and pulse conditions. The achieved system accuracy is reported, and its measurement capabilities are highlighted through the characterization of self-heating effects in LDMOS devices and silicon-on-glass VDMOS.
...
A low-cost, highly versatile, pulsed RF - pulsed I-V isothermal device characterization setup is presented. The realized setup combines a synthetic instrument high dynamic range pulsed network analyzer with pulsed I-V measurements. The resulting configuration facilitates very accurate characterization of low-power as well as high-power devices over a wide range of bias and pulse conditions. The achieved system accuracy is reported, and its measurement capabilities are highlighted through the characterization of self-heating effects in LDMOS devices and silicon-on-glass VDMOS.